Operating Instructions 

Incl. Declaration of Conformity 

Compact Pirani Gauge 

TPR 280 
TPR 281 

 

 

 

 

BG 805 178 BE / D     (2006-02) 

 
 
 

 

Product Identification 

In all communications with Pfeiffer Vacuum, please specify 
the information on the product nameplate. For convenient 
reference copy that information into the space provided 
below. 

 

Typ: 
No: 
F-No:
        V        W

Pfeiffer Vacuum, D-35614 Asslar

 

 
 

 

Validity 

This document applies to products with the following part 
numbers: 

TPR 280 

(W filament)

 

TPR 281 

(Ni filament)

 

 

PTR26950 
PTR26951 

PTR21950 

(DN 16 ISO-KF)

 

PTR21951 

(DN 16 CF-R)

 

PTR26960 
PTR26961 

PTR21960 

(DN 16 ISO-KF  long tube)

 

PTR21961 

(DN 16 CF-R 

long tube)

 

The part number (No) can be taken from the product name-
plate. 
If not indicated otherwise in the legends, the illustrations in 
this document correspond to gauges with DN 16 ISO-KF 
vacuum connections. They apply other vacuum connections 
by analogy. 
We reserve the right to make technical changes without prior 
notice. 
All dimensions in mm. 
 
 

 

Intended Use 

The Compact Pirani Gauges TPR 280 and TPR 281 have 
been designed for vacuum measurement of gases in the 
pressure range of 5×10

-4

 … 1000 mbar. 

The gauges must not be used for measuring flammable or 
combustible gases which react in air. 

They can be operated in connection with a Pfeiffer Vacuum 
controller for Compact Gauges or with another evaluation 
unit. 
 
 
 

 

Safety 

Symbols Used 

 

DANGER

 

Information on preventing any kind of physical injury. 

 

WARNING

 

Information on preventing extensive equipment and envi-
ronmental damage. 

 

Caution

 

Information on correct handling or use. Disregard can lead 
to malfunctions or minor equipment damage. 

 
 

 

Personnel Qualifications 

 

Skilled personnel

 

All work described in this document may only be carried 
out by persons who have suitable technical training and 
the necessary experience or who have been instructed by 
the end-user of the product. 

 
 

 

General Safety Instructions 

  Adhere to the applicable regulations and take the nec-

essary precautions for the process media used. 
Consider possible reactions between the materials and the 
process media. 
Consider possible reactions of the process media due to 
the heat generated by the product (e.g. explosions). 

  Adhere to the applicable regulationsand take the neces-

sary precautions for all work you are going to do and 
consider the safety instructions in this document. 

  Before beginning to work, find out whether any vacuum 

components are contaminated. Adhere to the relevant 
regulations and take the necessary precautions when 
handling contaminated parts. 

Communicate the safety instructions to all other users.

 

 
 

 

Liability and Warranty 

Pfeiffer Vacuum assumes no liability and the warranty be-
comes null and void if the end-user or third parties 

  disregard the information in this document 

  use the product in a non-conforming manner 

  make any kind of interventions (modifications, alterations 

etc.) on the product 

  use the product with accessories not listed in the product 

documentation. 

The end-user assumes the responsibility in conjunction with 
the process media used. 
Gauge failures due to contamination, as well as expendable 
parts (filament), are not covered by the warranty. 
 
 
 
 
 

 

Technical Data 

 

Measurement principle 

  thermal conductance according 

to Pirani 

Measurement range  
(air, O

2

, CO, N

2

 5×10

-4

 … 1000 mbar 

Accuracy (N

2

)  

 

1×10

-3

 … 100 mbar 

  ±15% of reading 

5×10

-4

 … 1×10

-3

 mbar    ±50% of reading 

100 … 1000 mbar 

  ±50% of reading 

Resolution 

  1% of reading 

Repeatability with air 

   

1×10

-3

 … 100 mbar 

  2% of reading 

 
Output signal 
(measurement signal) 

 

 

Voltage range 

VDC  0 … +9.0 

Measurement range 

VDC  +2.2 … +8.5 

Voltage vs. pressure 

 

logarithmic 
1.0 V/decade 

Error signal 

0 … +0.5  
(filament rupture) 

 
Output impedance 

 

2×4.7 

Minimum loaded 
impedance 

k

 

10, short-circuit proof 

Response time 

ms 

80 

 
Gauge identification 

  3.0 k

, referenced to supply 

common (voltage at pin 1 

5 V)

 
Adjustment 

  one tactile switch for ATM and 

HV adjustment 

 
Supply 

 

 

DANGER

 

The gauge may only be connected to power 
supplies, instruments or control devices that 
conform to the requirements of a grounded 
extra-low voltage (SELV-E according to 
EN 61010). The connection to the gauge has to 
be fused 

1)

 

 

 
Supply voltage 

 

 

At gauge 

VDC  +14 … +30 

Ripple V

pp

 

Current consumption 

mA 

<500  
(max. starting current) 

Power consumption 

Fuse required 

1)

 AT 

(slow) 

 
Electrical connection 

 

Hirschmann appliance 
connector, male,  
type GO 6, 6 poles 

Sensor cable 

 

5 poles plus shielding 

Cable length 

 

150 m (5×0.25 mm

2

200 m (5×0.34 mm

2

 
Grounding concept 

 

 "Electrical Connection" 

Vacuum connection to 
signal common 

 

connected via 1 M

 

(voltage difference <15 V) 

Supply common to 
signal common 

 conducted 

separately, 

for 

differencial measurement 

 
Materials exposed to 
vacuum 

 

DIN 1.4301, DIN 1.4305, 
DIN 1.4435, glass, Ni, 
NiFe 

Filament 

PTR26xxx 
PTR21xxx 

 

 

Ni 

 
Internal volume 

 

 

PTR26950, PTR21950 
PTR26951, PTR21951 

cm

3

 

cm

3

 

1.5 

1.5 

PTR26960, PTR21960

 

PTR26961, PTR21961 

cm

3

 

cm

3

 

10 

10 

Admissible pressure 

bar 
(abs.) 

10, limited to inert gases 

 

                                                                                 

1)

  Pfeiffer Vacuum controllers fulfill these requirements. 

(505)872-0037

idealvac.com

idealvac.com

Summary of Contents for TPR 280

Page 1: ...s are contaminated Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts Communicate the safety instructions to all other users Liability and Warranty Pfeiffer Vacuum assumes no liability and the warranty be comes null and void if the end user or third parties disregard the information in this document use the product in a non conforming manner make...

Page 2: ...n Vacuum Connection DANGER Caution overpressure in the vacuum system 1 bar Injury caused by released parts and harm caused by escaping process gases can result if clamps are opened while the vacuum system is pressurized Do not open any clamps while the vacuum sys tem is pressurized Use the type clamps which are suited to overpressure DANGER Caution overpressure in the vacuum system 2 5 bar KF conn...

Page 3: ...g and filter check that they are clean and replace them if necessary Deinstallation Activate the gauge and operate it at atmospheric pres sure for at least 10 minutes Press the button with a pin max ø1 1 mm and the ATM adjustment is carried out The gauge is adjusted to 1000 mbar 8 50 VDC by default By pressing the button 5 s the pressure value is increased towards 1200 mbar or by pressing it again...

Page 4: ...in accordance with the applicable regulations Organization company Address Post code place Phone Fax Email Name Company stamp Date and legally binding signature Copies Original for addressee 1 copy for accompanying documents 1 copy for file of sender Operating fluid s used Must be drained before shipping The service repair and or disposal of vacuum equipment and com ponents will only be carried ou...

Reviews: