![Pfeiffer TC400 Operating Instructions Manual Download Page 8](http://html.mh-extra.com/html/pfeiffer/tc400/tc400_operating-instructions-manual_1534853008.webp)
8
Connection "remote"
5
Connection "
remote
"
Remote control options are provided via the 26-pole D-Sub connector with the des-
ignation ”
remote
“ on the electronic drive unit.
Remove the remote plug from the TC 400 and connect a remote control unit. Pin
assignment of the connector according to table.
Shielded connectors and cables must be used.
5.1 Pin assignment
5.2 Operation via "
remote
" connection
+24 VDC* Output /
Pin 1
Inputs 2 - 6 and the connections to Pins 10, 13, 14 are activated by connecting them
with +24 VDC to Pin 1 (active high). They can also be activated via an external PLC.
The functions are deactivated by "PLC high level" and by "PLC low level".
• PLC high level: +13 V to +33 V
• PLC low level: -33 V to +7 V
• Ri: 7 k
Ω
Inputs
The digital inputs at connection "
remote
" are used to connect various functions of
the electronic drive unit. Functions are assigned to the inputs DI1 - DI2 ex factory.
These can be configured via interface RS485 and the Pfeiffer Vacuum parameter
set.
1
9
10
18
19
26
Pin
Function
Designation factory settings
1
+24 VDC output (V+)
Reference voltage for all digital in- and outputs
2
DI1
Enable venting; open: no; V+: yes
3
DI Motor pump
Drive motor; open: off; V+: on
4
DI Pumping station
Open: off; V+: on and error acknowledgement
5
DI Standby
Standby rotation speed; open: off; V+: on
6
DI2
Heating; open: off; V+: on
7
AI+ Rotation speed setting
mode
Set value in rotation speed setting mode;
2-10 VDC = 20-100% of the nominal rotation speed
8
DO1
Rotation speed switch point attained; GND:no;
V+: yes (I
max
= 50 mA/24 V)
9
DO2
GND: error; V+: no error (I
max
= 50 mA/24 V)
10
DI3
Sealing gas; open: off; V+: on
11
AI- Rotation speed setting
mode GND
Set value in rotation speed setting mode; GND
12
AO1
Actual rotation speed; 0-10 VDC is equivalent to 0-100%;
R
L
> 10 k
Ω
13
DI Error acknowledgement
Error acknowledgement: V+ pulse (min 500 ms)
14
DI Remote priority
Control via interface "remote"; open: off
V+: set and priority over other digital inputs
15
Relais 1
Connection to Pin 16 if relay 1 is inactive
16
Relais 1
Rotation speed switchpoint attained;
relay contact 1 (U
max
= 50 VDC; I
max
= 1 A)
17
Relais 1
Connection to Pin 16 if relay 1 is active
18
Relais 2
Connection to Pin 19 if relay 2 is inactive
19
Relais 2
No error; relay contact 2 (U
max
= 50 VDC; I
max
= 1 A)
20
Relais 2
Connection to Pin 19 if relay 2 is active
21
Relais 3
Connection to Pin 22 if relay 3 is inactive
22
Relais 3
Warning; relay contact 3 (U
max
= 50 VDC; I
max
= 1 A)
23
DO Remote priority
GND: off; V+: remote priority active
24
RS485 D+
according to specifications and Pfeiffer Vacuum protocol
25
RS485 D-
according to specifications and Pfeiffer Vacuum protocol
26
Ground (GND)
Reference ground for all digital inputs and all outputs