Duty to provide information on potential dangers
The product holder or user is obliged to make all operating personnel aware of dangers
posed by this product.
Every person who is involved in the installation, operation or maintenance of the product
must read, understand and adhere to the safety-related parts of this document.
Infringement of conformity due to modifications to the product
The Declaration of Conformity from the manufacturer is no longer valid if the operator
changes the original product or installs additional equipment.
● Following the installation into a system, the operator is required to check and re-evalu-
ate the conformity of the overall system in the context of the relevant European Direc-
tives, before commissioning that system.
General safety precautions
► Prevent process gases and their reaction and ancillary products that are harmful to health and the
environmental from escaping, or dispose of these according to the applicable regulations.
► Avoid contact with process gases that are harmful to health and if required, wear personal protec-
tive clothing.
► Check the media compatibility of parts carrying media.
► Observe all applicable safety and accident prevention regulations.
► Check that all safety measures are observed at regular intervals.
► Do not expose body parts to the vacuum.
► Never disconnect plug connections during operation.
► Never fill or operate the unit with cleaning agents or cleaning agent residues.
► Do not carry out your own conversions or modifications on the unit.
► Observe the unit protection degree prior to installation or operation in other environments.
► Provide suitable touch protection, if the surface temperature exceeds 70°C.
► Keep lines and cables away from hot surfaces.
► Before returning the vacuum pump, observe the notes in the chapter Service solutions.
2.4 Limits of use of the product
Installation location
Interiors, protected against the accumulation of
dust and weather influences, in a non-explosive,
dry environment
Protection degree IEC 60529
IP 20
Protection degree UL 50E
NEMA type 1
Installation altitude
max. 2000 m above sea level
1)
Ambient temperature
+5 °C to +40 °C
Permissible intake temperature of medium to
be pumped
+5 °C to +40 °C
Relative air humidity
80% at T ≤ 31 °C, to max. 50% at T ≤ 40 °C
Degree of contamination
2
Tbl. 3:
Permissible ambient conditions
2.5 Proper use
► Use the vacuum pump for vacuum generation only.
► Adhere to the installation, commissioning, operating, and maintenance instructions.
► Do not use any accessory parts other than those recommended by Pfeiffer Vacuum.
1)
at an installation location above 1000 meters above sea level, there is a risk of insufficient cooling. If necessa-
ry, take measures according to Directive DIN EN 61010.
Safety
13/54