18
Feature
Unit
TMH 262
TMU 262
TMTMH 262 P
TMU 262 P
Connection nominal diameter
Inlet
DN 100 ISO-K
DN 100 CF-F
DN 100 ISO-K
DN 100 CF-F
Outlet
DN 25 ISO-KF/G 1/4“
DN 25 ISO-KF/G
Venting connection
G 1/8”
G 1/8”
Nominal rotation speed
1/min
60 000
60 000
Standby rotation speed
1/min
40 000
40 000
Start-up time
min
3.5
3.5
Noise level
dB (A)
< 50
< 50
Final pressure, backing pump
mbar
< 5
< 5
Maximum permissible rotor temperature
°C
90
90
Permissible heat radiation power
W
8
8
Volume flow rate for
Nitrogen N2
l/s
210
210
Helium He
l/s
220
220
Hydrogen H2
l/s
190
190
Compression ratio for
N2
> 1 · 109
> 1 · 109
He
3 · 105
3 · 105
H2
1.3· 104
1.3· 104
Maximum fore-vacuum pressure for
N2
mbar
10
8.5
He
mbar
8
8
H2
mbar
4
4
Maximum gas throughput1)
With water cooling
N2
mbar l/s
4.7
3.2
He
mbar l/s
7.2
5.1
H2
mbar lIs
35.5
18.2
Maximum gas throughput at intake pressure of 0.1 mbar3)
N2
mbar l/s
7
6
He
mbar l/s
6
5
H2
mbar l/s
3.5
3
Vertex power characteristics line4)
A
W / Hz
120/1000
120/1000
B
W / Hz
120/1000
120/1000
C
W / Hz
120/1000
120/1000
D
W / Hz
120/1000
120/1000
Final pressure5)
With rotary vane pumps
mbar
< 5 · 10-10
< 5 · 10-10
With diaphragm pumps
mbar
< 1 · 10-8
< 1 · 10-8
Lubricant
TL 011
TL 011
Maximum cooling water
consumption with water at 15 ˚C6)
l/h
100
100
Cooling water temperature
˚C
5 - 25
5 - 25
Permissible ambient temperature
with air cooling
˚C
0 - 35
0 - 35
Heating power consumption
W
60
60
Weight
kg
4.8
6.8
4.8
6.8
Permissible magnetic field
mT
5.5
5.5
Operating voltage
VDC
24 ± 5%
24 ± 5%
Duration6)- / max. current
consumption
A
5.4 / 6.3
5.4 / 6.3
Duration6)- / max. power
W
130 / 150
130 / 150
Protection class7)
IP 30
IP 30
Relative humidity
%
5-85 non-condensing
5-85 non-condensing
9. Technical Data
1) Measured with a rotary vane pump 10 m3/h; higher throughputs with
reduced rotation speed.
2) Rotation speed of pump may drop below the nominal rotation speed.
3) For gas type characteristic lines please refer to section 4.4.
4) In accordance with German Industrial Standard 28428 the 48 hours
after baking out.
5) At maximum gas throughput.
6)
Protection class IP 54 is afforded for the Electronic Drive Unit TC 100