9
Modules
5.
Modules
Select the modules depending on which data you want to exchange with the drive unit.
➡
Thereby configure module modul 1 always first and use only once. As to restrictions
of the parameter values and parameter numbers please refer to the Operating
Manuals of the drive units used.
➡
Please also note the following:
Tabelle 5.0.1: Modules overview
Output data:
Contains data from the master (e.g. PLC) to the TIC.
Input data:
Contains data from the TIC to the master (e.g. PLC).
Configuration data:
Contains identification of the module.
Parametering data:
Contains additional module-specific information.
PLEASE NOTE
☞
–
A maximum of 18 modules can be used.
–
In the input data of the modules a maximum of 42 parameters may be present
[Pxxx]
.
–
In the configuration no other module may be present after the "fail_safe" mo-
dules.
–
In order to update the input data as rapidly as possible, please use only the mo-
dules you really require and do not use multiple parameter numbers.
–
Access to the functions in the modules can be influenced by external circuitry
of the drive unit used (refer to the Operating Manual of the drive unit).
#
Modules
Description
I-
O-
Data (Byte)
1
standard
Standard functions
4
3
2
power
Current and Voltage
4
0
3
ophrs_pump
Operating hours pump
4
0
4
ophrs_drv
Operating hours drive
4
0
5
TMS
TMS heating
1
1
6
venting
Venting
0
3
7
fixed_value_out
1 parameter value output
0
2
8
fixed_value_in
1 parameter value input
2
0
13
fail_safe
Behaviour if communication fails
0
0
14
acyclic access
acyclic access
12
8