13
Installation
Remove the connector plate and set the voltage selector switch "S" to the desi-
red voltage range.
Monitoring of the
operation conditions
The pump features the following functions for monitoring the pump components:
• Monitoring the temperature of the electronics
– Resetting by briefly removing the remote connector or by disconnecting from
the mains ("Off/On").
• Monitoring the internal motor winding temperature
– Resetting by briefly removing the remote connector or by disconnecting from
the mains ("Off/On"), followed by pump cooling.
• Surge current shut-down of electronics
– Resetting by briefly removing the remote connector or by disconnecting from
the mains ("Off/On").
5.5 Make remote connection
Fig. 8: Remote connector pin assignment M12 socket, 4-pin (A-coded)
2
+ 24 V*
2-1
Pump "On"
2-3
Temperature control mode "On"
2-4
Standby "On"
1
2 (+24 V)
1
3
4
2
3
4
Summary of Contents for Penta 10
Page 31: ...31...