14
Installation
ODK 2 with Adapter (with solenoid valve)
In order to avoid the pump being automatically ventilated via the operating fluid return
line after having been switched off, the magnetic valve (NC/normally closed) of the ODK
2 must be supplied with power at the same time as the engine.
Fig. 7:
Fit an ODK with a PK 005 998-T adapter on OME 25 M / OME 25 ML
Fit the connection plate with magnetic valve 300 at the gas ballast inlet so the elbow
is pointing upwards; pay attention to O-ring 301.
Insert outlet nozzle 210 in the elbow.
Fit operating fluid return hose 222 at both sides,
– keep the hose as short as possible and protect it from bending.
Tighten the union nuts of both screw fittings.
Fit coil 800 at the magnetic valve.
Establish a power supply at the coil, paying attention to the correct polarity and volt-
age.
222
210
300
800
700
33
101
100 102
103
40 41
145
301
PK 005 998 -T
OME
ODK 2
33
Drain screw for operating
fluid
40
Screw fitting
41
Reducer
100 Cover
101 Screws
102 Connection plate
103 Screws
145 O-ring
210 Outlet nozzle
222 Operating fluid return hose
300 Connection plate magnetic
valve
301 O-ring
700 Cable connection
800 Magnetic valve coil