Installation
13
Fit ODK 1 with adapter (without magnetic valve)
Fit connection plate ODK 1 at the gas ballast inlet so the elbow is pointing upwards;
pay attention to O-ring 301.
Insert outlet nozzle 210 in the elbow.
Fit operating fluid return hose 222 at both sides,
– keep the hose as short as possible and protect it from bending.
Tighten the union nuts of both screw fittings.
Fig. 6:
Fit ODK 1 with PK 005 998 -T OME 25 M / OME 25 ML adapter
NOTICE
Watch out for any accidental ventilation of the pump
After the ODK 1 has been fitted the pump is not sealed on the exhaust side. This results
in a vacuum chamber attached to the suction side being ventilated once the pump has
been switched off.
Make sure that on the side of a building a magnetic valve is installed on the suction
side.
33
Drain screw for operating
fluid
40
Screw fitting
41
Reducer
100 Cover
101 Screws
102 Connection plate
103 Screws
145 O-ring
200 ODK 1 connection plate
210 Outlet nozzle
222 Operating fluid return hose
301 O-ring
222
210
33
200
101
100 102
103
40 41
145
301
PK 005 998 -T
OME
ODK 1