1
2
3
1
Fig. 19:
Filling with lubricant
1 Filler screws
3
Fill level limiter
2 Drain screws
Filling with lubricant
1. Screw the fill level limiter back in.
2. Place a collection receptacle under each drain.
3. Fill the lubricant on both sides up to the maximum fill level.
– Once full, lubricant overfills the fill level limiter, and drips out of the drain hole.
4. Screw the filler and drain screws back in.
Maintenance