take notice with EBI-85A-TT
Storage temperature ..... -40°C bis +85°C
Measuring range ........... -85°C bis +85°C
Operating temperature . -85°C bis +85°C
Accuracy:
±0,3°C ........................... from -39°C to
+85°C
±2,0°C ........................... from -85°C to
-40°C
Life span of battery
typical 10 x 100 h ......... with -85°C
take notice with EBI-125A-EM-250/500-
F-1,6 (2K) and EBI-125A-R-250-F-1,5
Storage temperature ..... -40°C to +125°C
Measuring range ........... 0°C to +150°C
Operating temperature:
0°C ... +125°C
0°C ... +135°C max. 3h
0°C ... +150°C max. 1h
Measure accuracy
0°C ... 100°C ±0,3°C
100°C ... 140°C ±0,1°C
140°C ... 150°C ±0,3°C
Resolution ...................... 0.036 °C
Storage memory ........... approx. 6 000/
channel
Ex-accreditation EBI-85A, EBI-85A-Rx:
II 2G EEx ia IIC T4 (-20°C
≤
Ta
≤
+60°C)
II 2G EEx ia IIC T5 (-20°C
≤
Ta
≤
+50°C)
II 2G EEx ia IIC T6 (-20°C
≤
Ta
≤
+40°C)
á considerer EBI-85A-TT
Température de stockage
de -40°C á
+85°C
Etendue de mesure ....... de -85°C á +85°C
Température de fonctionnement
................................... de -85°C á +85°C
Exactitude:
±0,3°C ........................... de -39°C á +85°C
±2,0°C ........................... de -85°C á -40°C
Durée de vie de la pile:
typique 10 x 100 h ....... á -85°C
á considerer EBI-125A-EM-250/500-F-
1,6 (2K) et EBI-125A-R-250-F-1,5
Température de stockage
de -40°C á
+125°C
Etendue de mesure ....... de 0°C á +150°C
Température de fonctionnement:
de 0°C á +125°C
de 0°C á +135°C max. 3h
de 0°C á +150°C max. 1h
Exactitude:
de 0°C á 100°C ±0,3°C
de 100°C á 140°C ±0,1°C
de 140°C á 150°C ±0,3°C
Résolution .................... 0,036 °C
Mémoire ......................... env. 6.000/canal
Attestation de protection anti déflagrante
EBI-85A, EBI-85A-Rx
II 2G EEx ia IIC T4 (-20°C
≤
Ta
≤
+60°C)
II 2G EEx ia IIC T5 (-20°C
≤
Ta
≤
+50°C)
II 2G EEx ia IIC T6 (-20°C
≤
Ta
≤