V6.0
19
EN
8
Technical Data and Operation Conditions
Parameter
Description
Model
CardioSecur
Lead method
Bipolar, 3 lead
Reading quality
15-lead ECG with 4 electrodes
Pulse range
18 to 256 beats/minute
Precision
Band width: 0,05 to 40 Hz, sampling rate: 250 Hz
Type of protection
IP22
Storage
10,000 readings
Electrical safety
Application unit type BF
Power supply
through Smartphone
Temperature and humidity operation
+5 to +45°C, 10 to 95 %, non-condensing
Temperature and humidity storage
-20 to +45°C, 10 to 95 %, non-condensing
Air pressure for operation
700 - 1060 hPa
Electrodes
Disposible electrodes, Typ YD43-6-W or P-00-S/xx