ROHO MOSAIC Cushion Operation Manual

Cushion Instructions: 

C

USHION

 P

LACEMENT

 and 

H

AND

 C

HECK

Hand Check:

6. Perform a hand check to make sure the cushion is at the 

Proper Cushion Inflation Setting:
a.  Slide your hand between the cushion and the individual. 
b.  Slightly lift the individual’s leg and feel for their lowest 

bony prominence. 

c.  Lower their leg to a sitting position.

7. Adjust the cushion, if needed: 

a.  While keeping your hand under the individual’s lowest 

bony prominence, open the inflation valve. 

b.  Release air until you can barely move your fingertips. The 

cushion should be at the Proper Cushion Inflation Setting.

c.  Close the inflation valve.

Warnings: 

 

-

Make sure that there is at least ½ in. (1.5 cm) of air supporting the individual's lowest bony 
prominences.

 

-

After setting up the cushion the first time, check inflation frequently, at least once a day. Follow the 
Hand Check instructions above.

Close

Open

A) sink down into the cushion and 
B) “float” above the cushion base.

A

B

Proper Cushion Inflation Setting:

 

 = ½" - 1" (1.5 cm - 2.5 cm) 

recommended distance between 

the individual and the seating surface. 

Cushion Placement:

5. Place the cushion on the wheelchair with the cells right side up. 

 

-

The individual should sit in the wheelchair in their normal sitting position. 

 

-

Make sure that the cushion is centered underneath the individual. 

The inflation valve should be in the front, on the 
left side of the seated individual.

The seated figure shows product orientation. 

Left side

Right side

Summary of Contents for ROHO MOSAIC

Page 1: ...be given to the user of this product Operator Individual or Caregiver Before using this product read the instructions and save for future reference s h a p e f i t t i n g t e c h n o l o g y ROHO MOSAIC Cushion Operation Manual ...

Page 2: ...oting Storage Disposal and Recycling 6 Cleaning and Disinfecting 7 Product Specifications Limited Warranty 8 Español 9 Français 17 Deutsch 25 Italiano 33 Nederlands 41 Dansk 49 Svenska 57 Norsk 65 Suomi 73 Português 81 ROHO products may be covered by one or more U S and foreign patents and trademarks including ROHO shape fitting technology and MOSAIC 2017 ROHO Inc Rev 12 15 2017 ...

Page 3: ...r It will NOT support you in water If you are unable to perform any tasks described in this manual seek assistance by contacting your clinician equipment provider distributor or ROHO Inc DO NOT expose the cushion to any source of high heat including heating pads Exposure to high heat will accelerate aging and damage the product the benefits provided by the cushion may be reduced or eliminated incr...

Page 4: ...e carrying handle on the cover Using a pump cover or repair kit other than those provided by ROHO Inc may void the product warranty DO NOT modify the cushion or any components Doing so may result in damage to the product and will void the warranty Periodically check components for damage and replace if necessary See TROUBLESHOOTING To order replacement parts contact your equipment provider or ROHO...

Page 5: ...over the inflation valve Open the inflation valve b Inflate the cushion until all of the air cells feel firm c Close the inflation valve and remove the hand pump 2 Carefully follow the instructions for COVER PLACEMENT Cover Placement Note Keeping a cover on your cushion will help to protect the cushion components A cover will also help to keep the air cells together which may aid in slide board tr...

Page 6: ...inflation valve Warnings Make sure that there is at least in 1 5 cm of air supporting the individual s lowest bony prominences After setting up the cushion the first time check inflation frequently at least once a day Follow the Hand Check instructions above Close Open A sink down into the cushion and B float above the cushion base A B Proper Cushion Inflation Setting 1 1 5 cm 2 5 cm recommended d...

Page 7: ...dual and the wheelchair Allow the individual at least one hour to adjust to the cushion The Cushion Slides on the Wheelchair Make sure the non skid bottom of the cushion cover is facing away from the individual Use directional labels on the cushion and on the cover to orient the cushion Use hook and loop fasteners to further secure the cushion to the wheelchair The Cushion is Stiff If the cushion ...

Page 8: ...ning products that contain petroleum or organic solvents including acetone toluene Methyl Ethyl Ketone MEK naphtha dry cleaning fluid adhesive remover oil based lotions lanolin or phenolic based disinfectants ozone gas ultraviolet light cleaning methods To Clean the Cushion Remove the cover close valve s and place in a large sink Hand wash using mild liquid hand soap hand dishwashing liquid laundr...

Page 9: ...ure the cushion in the chair Refer to cover care label for material content Sizing all measurements are approximate W D W x D in W x D cm MOSAIC 1616 16 x 16 41 5 x 41 5 MOSAIC 1618 16 x 18 41 5 x 46 5 MOSAIC 1717 17 x 17 44 0 x 44 0 MOSAIC 1816 18 x 16 46 5 x 41 5 MOSAIC 1818 18 x 18 46 5 x 46 5 MOSAIC 2018 20 x 18 51 5 x 46 5 Cell Height Range 2 3 in 6 5 7 5 cm Width W and Depth D measured edge ...

Reviews: