37
Output stage
Output stage cable connections
The output stage is fitted with two screws.
Output stage R-net
Removal
1.
Switch off the main power switch on the control panel.
2.
Switch the main fuse to OFF. See page 40.
3.
Remove the chassis covers. See page 8.
4.
Disconnect the electrical connections from the output stage. Note their positions (see illustration
below).
Motor 1
Left
Motor 2
Right
Battery
Inhibit
External charging
socket
5.
Remove the output stage bracket by first loos-
ening the two screws (A) and then lifting the
bracket straight up while turning it upside
down. Then turn the bracket down and place it
on the left.
6.
Remove the output stage by unscrewing its
two mounting screws. See picture above.
Assembly
Assemble in the reverse order
.
Fit the output stage bracket with its lower section
inside the rear edge of the chassis (C). See the
illustration above.
The cable (B) that is fixed between the chassis and the out-
put stage bracket must not be disconnected.
The lower section of the output stage bracket is fitted inside
the rear edge of the chassis (C).
NB
A
. The output stage bracket is fitted using two
screws.
B
. The cable between the chassis and the output
stage bracket must not be disconnected.
C
. The lower section of the output stage bracket
is fitted inside the rear edge of the chassis.
B
A
C
Summary of Contents for C350
Page 1: ...Permobil C350 SERVICE MANUAL Power Wheelchair US ...
Page 2: ......
Page 62: ...62 Cabling overview A B C D E F 1 2 3 4 ...
Page 63: ...63 Cabling overview Cabling overview R net A B C D E F 5 6 7 8 ...
Page 64: ...64 Cabling overview A B C D E F 1 2 3 4 ...
Page 65: ...65 Cabling overview Cabling overview VR2 A B C D E F 5 6 7 8 ...
Page 67: ...67 ...
Page 68: ...Article no 205219 US 0 ...