Model 5210D User’s Manual
Doc. SE-MAN-050 REV 00
Page 6 of 25
©2017 Perma Pure LLC. All rights reserved. Specifications subject to change.
PHYSICAL DESCRIPTION
Single (Series) or Dual (Parallel) stream system
1 x 10” heat exchanger connected in series; actively cooled to the temperature set point
1 x 10” heat exchanger connected in series; passively cooled by direct contact with the heat sink
LCD temperature display
Dimensions: 14.5” x 11” x 10.5” HWD (37 x 28 x 26.5 cm)
Weight: 35 lbs (16 kg)
SPECIFICATIONS
Sample gas flow range
Up to 8 LPM (Up to 17 SCFH)
Avg. inlet dew point at rated flow
20% H
2
O (60°C DP) @ 5 LPM
Maximum cooling rate
440 BTU/hr (464 kJ/hr)
Maximum inlet sample temp.
400°F (205°C) for SS or glass impingers
280°F (138°C) for Kynar™ impingers
Maximum inlet pressure
15 psig (1 bar;760 mmHg)
Maximum pressure drop
<+1 in H2O
Ambient temperature range
33-104°F (0.6-40°C)
Outlet sample gas dew point
39°F (4°C)
Voltage
115 OR
230 VAC (Verify before energizing
equipment)
50/60 Hz
Power supply
600 Watt
Cooling down time
Less than 6 minutes
CONNECTIONS
Sample gas inlet: Kynar, 3/8” tube compression fitting
Sample gas outlet: Kynar, 1/4” tube compression fitting
Drain tubing connection: Kynar, ¼” barbed tube fitting
An automatic condensate drain, dual-head peristaltic drain pump is recommended
(not included)
for
water removal, Perma Pure part number 3KPB-003; size 17 tubing.
Always use factory supplied fittings; Perma Pure cannot warrant against damage to the Peltier elements or heat
exchangers if our fittings are not used.