58 . TGA 8000 Installation and Hardware Guide
TGA 8000 Furnace Maintenance
Standard furnace maintenance consists of the following procedures:
•
•
•
Removing the Furnace
WARNING:
Before performing this procedure, shut down the system and remove
the line power. Make sure the furnace has cooled sufficiently so that you do not
burn yourself when removing the furnace components.
To perform this procedure the furnace must be in the lowered position.
AVERTISSEMENT:
Avant d'effectuer cette procédure, arrêtez le système et
retirez l'alimentation secteur. Assurez-vous que le four a suffisamment refroidi
pour ne pas vous brûler lors de la dépose des composants de la fournaise.
Pour exécuter cette procédure, le four doit être en position abaissée.
1.
Disconnect the furnace power cord from its connector located behind the left panel
of the TGA. The furnace cable is screwed into the connector. To disconnect unscrew
and gently pull forward on the connector.