60 . TGA 8000 Installation and Hardware Guide
If you plan to use the thermocouple from the removed furnace on the new furnace, remove
it as described in “Replacing a Thermocouple.”
Replacing the Furnace
1.
Insert thermocouple into furnace base following the thermocouple replacement
procedure
2.
Place the furnace on the furnace base. Note that there is a locating pin on the
mounting assembly that fits into the hole on the bottom of the furnace to assist
alignment.
3.
Place the furnace assembly onto the mounting assembly.
4.
Place the knurled furnace locking nut over the furnace.
5.
Place the clip underneath the knurled furnace locking nut.
6.
Straighten the furnace and tighten the knurled nut.
7.
Slide the platinum furnace shield over the furnace.
8.
Place the spring over the furnace and until it sits in the top of the metal furnace
base. Make sure that the top of the spring is straight.