30 . Spotlight 150 User's Guide
Infrared Optics
The upper cassegrain is used for both visible light and infrared radiation. For this reason,
when you adjust the sample position so that the visible image of the sample is in focus, the
sample is also correctly positioned for collecting infrared spectra. Similarly, when you adjust
the aperture so that the required area of the sample is isolated visually, you have also
isolated the area of the sample from which the IR spectrum is to be collected.
IR optics in transmittance
IR optics in transmittance differs from when viewing a sample in transmittance as follows
•
Instead of receiving light from the illuminator, light from the spectrometer is reflected
off the toroid onto the lower dichroic mirror which sends it through the lower
cassegrain.
•
The upper dichroic mirror reflects the beam onto the detector cassegrain.
•
The detector cassegrain focuses the beam on to the detector.
Figure 9 Path of the infrared beam for collecting an image in transmittance
NOTE:
Figure 9, above, illustrates a Spotlight 150 with an MCT detector. Other detector
types may look different, but the optical arrangement is the same in all cases.
Dichroic mirror
Variable
aperture
Upper
cassegrain
Sample position
Lower
cassegrain
Lower dichroic
mirror
Detector cassegrain
Detecto
r
Source toroid
Summary of Contents for Spotlight 150
Page 1: ...Spotlight 150 User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 11: ...Warnings and Safety Information ...
Page 23: ...Overview of the Spotlight 150 ...
Page 32: ...32 Spotlight 150 User s Guide ...
Page 33: ...Getting Ready to Use the Spotlight 150 ...
Page 45: ...Preparing Samples ...
Page 58: ...58 Spotlight 150 User s Guide ...
Page 59: ...Techniques for Collecting Spectra ...
Page 94: ...Maintenance ...
Page 102: ...Appendices ...