The following table details the functionality of the pins. To connect a third-party external 

detector, present an analog differential signal to pins 10 and 11. 

 
Pin 

number 

Function 

Notes 

Accessory Bus Reset 

System reset, normally 5 V but 0 V pulse on 

spectrometer power up. Output. 

0 VL 

0 V return for digital circuitry. 

Accessory Bus Clock 

30 7200 Hz TTL compatible signal for Accessory 

Bus microcontroller clock drive. Do not pull 

more than ±2 mA from this pin. Output. 

Accessory Bus Abort 

Accessory abort to prematurely finish Accessory 

Bus command. Output. 

Accessory Bus serial transmit 

(from spectrometer to 

external) 

TTL compatible serial signal running at 4800, n, 

8, 1. Output. 

Accessory Bus serial receive 

(from external to 

spectrometer) 

TTL compatible serial signal running at 4800, n, 

8, 1. Input. 

0 VL 

0 V return for digital circuitry. 

5 VL 

5 V supply for digital circuitry. Do not pull more 

than 1 A from this pin. Output. 

0 VA 

0 V analog return for analog circuitry. 

10 

 

External differential positive analog detector 

signal. ±11 V full scale. Input. See Note 1. 

11 

EXTDET− 

External differential negative analog detector 

signal. ±11 V full scale. Input. See Note 1. 

12 

0 VA 

0 V analog return for analog circuitry. 

13 

+12 VA 

+12 V analog for any analog circuitry. Do not 

pull more than 100 mA from this pin. 

14 

−12 VA 

−12 V analog for any analog circuitry. Do not 

sink more than 100 mA into this pin. 

15 

TRIG_IN (NOT 

IMPLEMENTED, LEAVE 

DISCONNECTED) 

Trigger in. Not implemented at the present 

time. Input. 

 

NOTE:  

The differential input signal has a maximum positive value when  is at +11 V and EXTDET− is at −11 V. It 

has a maximum negative value when  is at −11V and EXTDET− is at +11 V. The input impedance is 100 

KΩ on each input with reference to 0 VA. 

Summary of Contents for Spectrum 100 Series

Page 1: ...pre amplifier and detector of your choice Figure 1 Handhold with external accessory bracket fitted This document describes how to Fit a different window in the right hand external beam port Connect a...

Page 2: ...netic interference and electromagnetic compatibility The Right hand External Beam Option is supplied for use as an experimental test bed The user configurable nature of this product means that PerkinE...

Page 3: ...nd leave this port open to atmosphere If the spectrometer is fitted with KBr optics do not exchange the window if the humidity is 45 relative humidity The humidity sensitive optics will be damaged if...

Page 4: ...d instructions refer to the Advanced Maintenance section of the Spectrum 100 Series User s Guide L1050021 or Spectrum 400 Series User s Guide L1050056 These are distributed as pdf files on the Spectru...

Page 5: ...cure the main cover 9 Plug in and switch on the instrument Fitting a Third party Pre amplifier and Detector to the Instrument The EXT R socket on the instrument is a 15 Way High Density D type The con...

Page 6: ...nal running at 4800 n 8 1 Input 7 0 VL 0 V return for digital circuitry 8 5 VL 5 V supply for digital circuitry Do not pull more than 1 A from this pin Output 9 0 VA 0 V analog return for analog circu...

Page 7: ...guration file for use with a third party external detector NOTE It is not necessary to power the instrument to complete this procedure 1 Browse to C pel_apps bin The instrument configuration file is n...

Page 8: ...nfiguration file is opened for editing 6 Right click in the external accessory area to the right of the beampath diagram The Add Port dialog is displayed 7 Select Right Mid in the Select Site pane sel...

Page 9: ...t Type pane and then click Install For information on the options available to you see Choosing the correct detector type on page 10 The Add Detector dialog closes and becomes 11 Right click The Add m...

Page 10: ...ed Custom MIR 15000 370 15 secs 0 2 cm s Custom NIR 15000 1250 15 secs 0 2 cm s Custom FIR 720 30 15 secs 1 0 cm s See Adjusting the default Custom detector settings on page 10 for details of how to a...

Page 11: ...rum online help Stabilization time The time required in seconds for the detector to settle after the infrared beam is directed onto it that is after the beampath is changed to use the external detecto...

Page 12: ...original internal detector configuration file as C cfg NOTE If a backup of the original configuration file was not made uninstall the instrument and then re install it using the configuration disk su...

Page 13: ...g you that the detector may be damaged by the NIR beam You should only select the Yes option to continue if an attenuator has been fitted in the beam or you are confident that the detector can withsta...

Page 14: ...lect the Setup menu and then click Instrument 2 Select the Setup Instrument BeamPath tab 3 Select Right external as the Beam Location from the drop down list The external detector is selected NOTE The...

Page 15: ...ed The Display tab is displayed OR If you have Spectrum Express software switch to Monitor by clicking on the Measurement toolbar and then click Energy to display the signal detected The Live tab is d...

Reviews: