![PerkinElmer OMNI International Bead Ruptor 96 Manual Download Page 9](http://html1.mh-extra.com/html/perkinelmer/omni-international-bead-ruptor-96/omni-international-bead-ruptor-96_manual_1530222009.webp)
Replacing the Fuse
Troubleshooting
Troubleshooting
Error
Possible Solution
Power Switch is not illuminated
when the Bead Ruptor 96 is on.
1. Check the power supply
2. Check fuses and replace if necessary
3. Call for service
Display screen is not working
Call for service
“LID” appears on display
Indicates that the hood is open when the start button
was pressed. Close hood and clear message by pressing
the Start/Stop button. Press Start/Stop to begin cycle.
Message “Er1” appears on
display screen
Motor has over heated. Turn power off and allow motor
to cool down.
Message “Er2” appears on
display screen
The Bead Ruptor 96 is unable to reach the set vibrational
frequency. Check Milling Jars for obstructions. Switch
Power OFF and ON to clear error message.
Message “Er3” appears on
display screen
Hood was opened during operation. The power must be
turned OFF and then ON to clear error message before
operation can continue.
DO NOT: attempt to service the Bead Ruptor 96 in a manner other than
those discussed in this manual. For any issue that is unsuccessfully correct-
ed using this guide, please contact your authorized dealer or call
Omni International at 1-800-776-4431.
Replacing the Fuse
The following fuses are required for the Bead Ruptor 96:
• 2 x 2AT 250V (230V)
• 2 x 5AT 250V (115V)
• Unplug power cord from the power cord port (2).
• By pressing the locking device on the bottom side of the fuse compartment (1), fuse
holder is released and you can pull it out.
• Replace fuses.
• Insert fuse holder and push it, until it locks.
1. Fuse Compartment
2. Power Cord Port
The following fuses are required for the Bead Ruptor
TM
96 Homogenizer:
• 2 x 2AT 250V (230V)
• 2 x 5AT 250V (115V)
• Unplug power cord from the power cord port (2).
• By pressing the locking device on the bottom side of the fuse compartment (1), fuse
holder is released and you can pull it out.
• Replace fuses.
• Insert fuse holder and push it, until it locks.
DO NOT: attempt to service the Bead Ruptor
TM
96 Homogenizer in a manner other
than those discussed in this manual. For any issue that is unsuccessfully corrected
using this guide, please contact your authorized dealer or call
OMNI International, a PerkinElmer company at 1-800-776-4431.
Error
Possible Solution
Power Switch is not illuminated when the Bead
RuptorTM 96 Homogenizer is on.
1. Check the power supply
2. Check fuses and replace if necessary
3. Call for service
Display screen is not working
Call for service
“LID” appears on display
Indicates that the lid is open when the start
button was pressed. Close lid and clear mes-
sage by pressing the Start/Stop button. Press
Start/Stop to begin cycle.
Message “Er1” appears on display screen
Motor has over heated. Turn power off and
allow motor to cool down.
Message “Er2” appears on display screen
The Bead Ruptor
TM
96 Homogenizer is unable
to reach the set vibrational frequency. Check
Milling Jars for obstructions. Switch Power
OFF and ON to clear error message.
Message “Er3” appears on display screen
Lid was opened during operation. The power
must be turned OFF and then ON to clear
error message before operation can continue.