PerkinElmer OMNI International Bead Ruptor 96 Manual Download Page 7

Inserting Well Plate Adapters

• 

Bead Ruptor

TM

 96 Homogenizer

 well plate adapters can hold 1 X 2 mL deep well plate,

  2 X 1 mL deep well plates or 3 X standard well plates per processing arm.
• 96 well 2D barcode storage tube adapters can hold 1 x 1.1 ml or 1 x 1.4 ml plates per

  processing arm.
• Place samples and beads in the well plate.
• Fully seal the well plate with well plate mat. Care must be taken to ensure the well
  plate is properly sealed. A silicon mat is recommended.
• If processing multiple well plates, stack the well plates bottom to top.
• Place the well plate adapters on the top and bottom of the well plate stack. Ensure
  the adapter extensions are aligned.
• Lift the locking pin (B) upwards from the groove and turn it 90° in any direction

  to unlock.
• Turn the knob (C ) counter clockwise all the way to maximize clamping range.
• Turn locking pin 90° and allow it to reengage in locked position.

• Place the well plate adapter into the milling jar arm and seat it into centering points (A).

  Gently turn knob (C ) clockwise until the well plate adapter fits firmly inside the holder,
  free of play.
• To remove the well plates, lift the locking pin (B) and turn it 90° to unlock and turn
  the knob (C) counter clockwise to loosen. If locking pin is difficult to release, turn
  knob (C) slightly clockwise to release locking pin tension.

WARNING: Do not force locking pin (B) open. Turn knob (C ) clockwise to release.
WARNING: Well plates and adapters must be inserted and aligned with centering 

points (A).  If this is not done the well plates may be ejected during processing.
WARNING: Milling arms must be balanced prior to operation. You must always 

operate with two well plate adapters. One in each arm.
WARNING: The Bead Ruptor

TM

 96 only accepts SPS format plates less than 49.25 mm high.

Summary of Contents for OMNI International Bead Ruptor 96

Page 1: ...Bead RuptorTM 96 Well Plate Homogenizer...

Page 2: ...EXCEED THE CONTRACT PRICE FOR THE GOODS FOR WHICH LIABILITY IS CLAIMED This product has been engineered for safety however basic safety precautions and common sense must always be demonstrated when u...

Page 3: ...200W Fuses 2 x 2AT 250V 230V 2 x 5AT 250V 115V Vibrational frequency regulation Digital from 3 to 30 Hz 180 1800 min 1 in 0 1 Hz steps Timer 5 sec 99 min timer HOLD function in 1 sec steps below 10 mi...

Page 4: ...e Bead RuptorTM 96 Homogenizer frequency selection is active Frequency is defined as number of wave cycles per second 1 s Measured in Hz CONTROL KNOB rotate push rotate right Or left to change the Tim...

Page 5: ...A A B B C C A Centering Points B Locking Pin C Knob Milling Jars O Rings Jar Wrenches 96 Well Plate And 2D Barcode 96 Well Storage Tube Adapters Milling Jar and Rack Holders...

Page 6: ...t 90 to unlock and turn the knob C counter clockwise to loosen If locking pin is difficult to release turn knob C slightly clockwise to release locking pin tension WARNING Always include an O Ring whe...

Page 7: ...r clockwise all the way to maximize clamping range Turn locking pin 90 and allow it to reengage in locked position Place the well plate adapter into the milling jar arm and seat it into centering poin...

Page 8: ...when working with liquid nitrogen Liquid nitrogen has a temperature of 196 C and will cause severe burns and frostbite with skin or eye contact Milling Jars Order Number Stainless Steel 25 mL set of t...

Page 9: ...d push it until it locks 1 Fuse Compartment 2 Power Cord Port The following fuses are required for the Bead RuptorTM 96 Homogenizer 2 x 2AT 250V 230V 2 x 5AT 250V 115V Unplug power cord from the power...

Page 10: ...PN 03 296 Rev G 935 Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com...

Reviews: