09931262E
12
b.
Experiment: SBW is set 2.0 nm automatically. Setup other experiment parameters.
c.
Lamp: Single Front is automatically selected. If not, select Single Front and setup other
experiment parameters.
d.
Setup the Quantification parameters. For more information see the
Lambda 365 UV Express
Software Users Guide, the section on Scanning Quantification.
5.
Select Apply and OK after set-up of parameters.
6.
Dip the probe in the blank solution and select the Blank icon. Absorbance of selected wavelength
is changed about 0 Abs.
NOTE:
Before measurement, make sure that that bubble does not exist at the mirror side of the probe.
NOTE:
When sample is measured, any light from outside of the bottled should be blocked as it affects the
measurement result.
7.
Clean the probe and Dip the probe in the Standard solution. Select the Standard icon. Measure
standard solutions in order and the calibration curve will be created.
8.
Clean the probe and dip the probe in the sample solution. Select the Sample icon.
9.
Input sample name and select OK.
10.
The spectrum and result will be displayed. Save or print results as desired.