Unpacking and Installation . 41
Other connectors
The communication ports for peripheral devices are shown in Figure 13.
Figure 13 Communication ports
Legends and icons that identify the function of each of the ports are printed on the hinge
molding directly above the port.
CAUTION
Do not attempt to connect a monitor to either the EXT.R port or the
MICROSCOPE port or you will cause serious damage to the instrument
when it is switched on.
Connector details
Icon
Description
Connector
Type
Voltages Maximum
Currents
10101
Serial port. Connects a PC to the
instrument using an RS232
interface cable. This connector is
only used for service diagnostics.
9-way
D-type
+12 V
<100 mA
−12 V
<100 mA
General Purpose Optical Bench
(GPOB). Allows synchronization
between the instrument and an
external accessory.
26-way
high density
D-type
+5 V
<50 mA
Spt.Lt
Spotlight. Used to connect the
instrument to a Spotlight
Imaging System.
11-way mixed
D-type
+5 V
<100 mA
±12 V
50 mA
Sample area purge
Instrument purge
10101
Spt. Lt.
EXT.R
Summary of Contents for Frontier IR
Page 1: ...FRONTIER IR DUAL RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 Frontier IR Dual Range Systems User s Guide...
Page 11: ...Warnings and Safety Information...
Page 23: ...An Overview of the Frontier IR Dual Range Spectrometers...
Page 34: ...34 Frontier IR Dual Range Systems User s Guide...
Page 35: ...Unpacking and Installation...
Page 45: ...Using the Spectrometer with Spectrum...
Page 52: ...52 Frontier IR Dual Range Systems User s Guide Figure 21 Atmospheric suppression...
Page 57: ...Routine Maintenance...
Page 78: ...78 Frontier IR Dual Range Systems User s Guide...
Page 79: ...Advanced Maintenance...
Page 94: ...94 Frontier IR Dual Range Systems User s Guide...
Page 95: ...Appendices...