7 CORRECTION MODE
7-1
Changing the Target Constant
Changing the Target Constant can be performed only when the Reflector sheet
and Prism Constant settings are “INPUT” in Initial Setting 1.
Example: Prism Constant - 25mm setting
Press [F4] [CORR] in mode B.
(If the instrument is in mode A, press [F5]
[MODE] to enter mode B.)
(SHEET CONST : Reflector sheet constant)
Press the [F5] [SELECT] to enable the Prism
Constant to be changed.
Clear the exiting values by pressing [CLEAR] key.
Input 25 by pressing the numeric keys.
46
7 CORRECTION MODE
SELECT
CORRECTION
1.PRISM CONST : -30mm
2.SHEET CONST
0mm
3.TEMP
15
°
C
4.PRESS
1013hPa
5.ppm
0ppm
CORRECTION
1.PRISM CONST : -30mm
2.SHEET CONST 0mm
3.TEMP
15
°
C
4.PRESS
1013hPa
5.ppm
0ppm
CLEAR
PENTAX R-300 (p20-52) 10/30/02 11:08 AM Page 46