Calibration
Calibration and adjustment in the traditional
sense are not necessary as long as the instru-
ment is operating in disciplined mode. Then the
internal rubidium timebase is continuously
monitored and forced to follow the cesium
clocks of the navigational satellites. These
clocks are traceable to, for instance, NIST and
USNO. Thus the normal aging characteristics
of the internal timebase will be concealed.
However, if the device is operating in hold-over
mode for long periods, the aging characteristics
are no longer concealed. Then you may need to
reset the timebase from time to time by letting
the device operate in disciplined mode for at
least 48 hours. See also next paragraph.
n
Calibration Intervals
Normally it is quite adequate to check the fre-
quency deviation between one of the reference
outputs and a corresponding, independent, ce-
sium-controlled reference source once a year,
using the procedure described below. Shorter
intervals may be necessary if the main operat-
ing mode is
hold-over
, and the allowed devia-
tion is less than the specified annual aging. See
the timebase specifications to collect the data
for these calculations.
n
Equipment
–
The DUT (Device Under Test).
–
Timer/Counter Pendulum CNT-90.
–
A GPS-controlled reference frequency
source capable of generating both 2.048
MHz and alternatively 1, 5 or 10 MHz at
an amplitude between 0.1 and 5 V
rms
in
50
W
. It should also have a traceable cali-
bration history. The Pendulum GPS-89
with Option 72 is apt to the task.
–
BNC cables of suitable lenghts.
Note:
There are several ways to perform
this checkup, depending on available
equipment. The setup and procedure
described in the next paragraphs are
based on the list above.
n
Setup
–
Connect the reference source (1, 5 or
10 MHz) to the
EXT REF
input on the
counter.
–
Connect one of the 2.048 MHz outputs on
the front panel of the DUT to
Input A
on
the counter.
–
Connect the 2.048 MHz reference source
to
Input B
on the counter.
–
Find a good antenna position, and connect
the antenna to the DUT. Power up the
DUT and the reference source at least 24
hours before starting the verification pro-
cedure. Make sure the GPS contact is es-
tablished and the system is working within
15 minutes after the DUT was switched
on.
–
Power up the timer/counter at least one
hour before starting the verification proce-
dure.
n
Procedure
Note:
Each key depression is not described
here. See the Operators Manual for
the timer/counter if you need more
information.
–
Recall the default settings of the counter.
–
Set the trigger level to
Man 0.0 V
on both
Input A
and
Input B
.
–
Set the measuring time to 10 s.
–
Check that the frequency is 2.048 MHz ±
0.01 Hz.
–
Change measurement function to
Time In-
terval A to B
.
6-2 Calibration
Preventive Maintenance
Summary of Contents for GPS-12
Page 1: ...GPS 12 GPS 12R GPS 12R HS GPS 12RG GPS GLONASS GPS Frequency Standards User s Manual...
Page 2: ...II 4031 600 12001 Rev 05 May 2017 2017 Pendulum Instruments Altaria Services...
Page 5: ...Chapter 1 Preface...
Page 8: ...This page is intentionally left blank 1 4 Preface...
Page 9: ...Chapter 2 Preparation for Use...
Page 17: ...Chapter 3 Using the Controls...
Page 24: ...This page is intentionally left blank 3 8 User Interface Using the Controls...
Page 25: ...Chapter 4 GPS 12 Monitor...
Page 28: ...This page is intentionally left blank 4 4 Main Features GPS 12 Monitor...
Page 29: ...Chapter 5 Performance Check...
Page 35: ...Chapter 6 Preventive Maintenance...
Page 38: ...This page is intentionally left blank 6 4 Battery Replacement Preventive Maintenance...
Page 39: ...Chapter 7 Specifications...
Page 48: ...This page is intentionally left blank 7 10 Ordering Information Specifications...
Page 49: ...Chapter 8 Appendix...
Page 57: ...Chapter 9 Index...
Page 61: ...Chapter 10 Service...