background image

RX™ 22

/22XT

 Replacement Kit Instructions

Peavey Electronics Corporation • 5022 Hartley Peavey Drive • Meridian, MS 39305

Phone: (601) 483-5365 • Fax: (601) 486-1278 • www.peavey.com •

 80302569 • ©2007 Printed in the U.S.A.

1)

Remove the diaphragm assembly from the driver by removing the six screws in the top cover.

2)

Remove the old ferrofluid by inserting a strip of filter paper into the gap, allowing it to sit for one (1) minute. Remove the strip 

and 

repeat the process using a fresh piece of paper until no ferrofluid remains in the gap. If you run out of filter paper, any absorbent 

paper can be used to complete the job.

3)

Inspect the magnetic air gap for the presence of any debris. If necessary, clean the gap with a piece of folded masking tape 

(sticky side out).

4)

Using the notch in the side of the packet as a locating guide, unscrew the fluid-filled bottle. Pour out the contents of the bottle 

into the magnet structure gap.
Note:

 

It is not necessary to rotate

 

the magnet;

 

the ferrofluid is self-distributing. Attempting to dispense the remaining ferrofluid 

may over fill the air gap, which has no benefit.

Note: Remove and discard the enclosed Spacer Shim if using the RX22.

5)

Replace the diaphragm assembly and firmly tighten the six screws. Word of advice: When tightening the screws try to tighten 
them

 

down evenly, the same amount as you go around the cover. This will help eliminate any mechanical distortion in the 

diaphragm. Thank you for choosing Peavey as your musical instrument supplier.

Instructions for Use

 with RX 22

:

Before continuing, make sure this package contains the following:

One (1) ferrofluid filled packet

This instruction sheet/ Material Safety Data Sheet

R

22 replacement diaphragm

GAP

LOADING

PLUG

SCREW HOLE

FRONT PLATE

LOCATING PIN

HOLE

Procedure:

Instructions for 

22XT+™ Diaphragm

Replacement Kit

These instructions are broken into three sections:

1. Driver Inspection, 2. Diaphragm Replacement

Procedure, 3. Failure Analysis. In most cases, following

this order will assure your success.

*Interchangeable with all Model 22A and Model 22

Drivers.

DRIVER INSPECTION

1. Driver Dead: With an Ohm meter (or in a pinch, a

flashlight battery), determine if the driver is open circuited

by measuring at the driver terminals. D.C.

resistance should be in the range of 4.25 Ohms to 4.75

Ohms. Unless you have a precision-calibrated meter

or a bridge, consider the driver to be okay if you're in

the ballpark. If using a battery, no “click” when connecting

and disconnecting the battery means the driver

is open.

2. Driver Noisy or Distorted: If you have a sine wave

oscillator, sweep the driver with 4 Volts from 500 Hz to

2,000 Hz. Any buzzing or distortion will indicate that

the diaphragm must be removed. An intermittent noise

could be a metal particle in the gap. Check the gap

carefully and clean before you replace the diaphragm

to rule out this possibility. A flashlight is a great help. If

an oscilloscope is available, look at the waveform at

the driver to rule out an electronic malfunction if high

distortion is your symptom.

DIAPHRAGM REPLACEMENT PROCEDURE

(see Figures A and B on reverse side)

1. Clean the front plate and environment of all metal
particles.

2. Remove the six 8-32 screws holding the diaphragm

assembly to the front plate. These will be replaced with

a longer screw, included (71500875).

3. Remove the failed diaphragm assembly to a safe

place to prevent damage so that we can inspect it.

(You might want to look closely at it, too.) Please return

the failed diaphragm assembly to your dealer or

directly to Peavey using the box in which the new

assembly arrived. Describe the operating conditions

surrounding the failure with as much detail as possible.

This information will help immeasurably in improving

the performance of our products.

4. Fold a piece of 3/4" wide masking tape into a triangle

shape with the adhesive side out. Using a flashlight

to see into the gap in the magnet structure, slide tape

around the entire gap to clean out all metal chips and

dirt.

6

. Make certain the rubber gasket is properly seated

on the new cover assembly, exposing the entire aluminum

ring. The ring must sit flat on the front plate

when in place.

7

. By eye, line up the locating pins on the cover with

the untapped holes on the front plate. These holes are

easy to locate since they are between tapped holes for

the 8-32 screws. If the gasket will not stay on the

cover, tilt the driver past vertical and let gravity help

you.

8

. Gently push the diaphragm assembly onto the driver,

making certain that the locating pins are going into

their holes.

9

. With the cover in place, carefully check the gasket

to be sure it hasn't slipped under the ring. (If it has,

start over or use tweezers to reposition the gasket.)

10

. Now that you have assured yourself that the

diaphragm assembly is seated solidly on the front

plate, put the new, longer 8-32 screws into the holes

and tighten. A tightening sequence is not necessary.

GASKET

LOCATING PIN 

SUSPENSION

DIAPHRAGM

VOICE COIL

Cover Assembly

FAILURE ANALYSIS

You may be able to learn why your driver failed by

inspecting the failed diaphragm assembly. We powertest

and rigorously analyze each diaphragm assembly.

Consequently, there is more than an even chance

that the diaphragm assembly failed because of too

much power input.

1. The voice coil is copper in color initially and turns

dark brown, even black, under hard use. This is due to

a color change in the epoxy adhesive on the wire

(occurs at 300° F).

Any charring on the wire is a sign of too much power.

If some turns have come off the coil and the wire is

shiny, most likely the coil was rubbing on the gap walls

before failure.

2. Diaphragm Broken, Cracked or Dented: When

the driver is driven with high amplitude, low frequency

signals at or below the low frequency limit of 500 Hz,

the diaphragm bottoms on the loading plug, causing

the above mentioned damage; also, input power

above the design limit within the passband of the driver

can cause this damage.

3. Determining Power Input to the Driver: This is an

extremely difficult proposition with program material

even with very sophisticated measuring instruments.

The formula that applies is power=(voltage x voltage)

/ impedance. Assume the impedance listed on the

diaphragm assembly label and use either a calibrated

oscilloscope or Voltmeter to determine Voltage. The

driver is capable of withstanding many times rated

power for short duration but average long term power

(measuring RMS Voltage) is 40 Watts. If your amplifier

is oscillating or your P.A. is operated in feedback,

unusually high Voltage could be applied to the driver

causing failure.

We appreciate your interest in our product; if you

require further assistance, please contact your dealer

or our factory.

Figure A

Figure B

Note:  The 

XT 

spacer

  shim

  (gasket)  is  only  used 

on  the  22XT+  and  all  other  older  22  drivers. 

RX22 does not use the spacer.

 

Note: While a necessary component for the RX 

22, the ferrofluid is optional for the 22xt+ kit. 

Using the ferrofluid, however, would be an 

added benefit for the 22xt+ kit, by reducing the 

heating of the voice coil, improving the power 

rating and smoothing the frequency response in 

the mid region of the driver.

Logo referenced in Directive 2002/96/EC Annex IV

(OJ(L)37/38,13.02.03 and defined in EN 50419: 2005

The bar is the symbol for marking of new waste and 

is applied only to equipment manufactured after

13 August 2005

*see Ferrofluid Safety Information on Reverse side of this 

sheet.

XT 

Spacer 

Shim (for use with 22XT only)

8-32 screws (for use with 22XT only)

5. 

Install 

the included XT spacer

 shim

.

Reviews: