9.5. Remote Control Mode (P 1530 / 1560)
There are two methods for remote control of current and voltage adjustment. Both methods require
current remote control part to be set up in order for remote control mode to be functional, otherwise
unit will be in CC mode all the time.
Method A:
Using two external variable DC voltage sources.
PIN
Function
Remarks
1
Internal DC +5V
Less the 50mA
2
Voltage Adjust
0 ~ 5 V
3
Current Adjust
0 ~5 V
4
Ground
5
Output OFF
Short to Ground
6
N.A.
7
N.A.
8
N.A.
Check the output voltage range of the power supply by varying the external voltage source.
Short circuit the main output with 8AWG (8,35mm2) wire to check the display for CC setting varying
the external voltage source.
Method B:
Using two 0-5K Ohm variable resistors
Remark: variable resistors 5KOhm
Check the output voltage range of the power supply by adjusting the 5Kohm variable resistor. Short
circuit the main output with 8AWG (8,35mm2) wire to check the display for CC setting by adjusting the
variable resistor.
-38-
PIN
Function
Remarks
1
Internal DC +5V
Resistor end
2
Voltage Adjust
Variable part of resistor V
3
Current Adjust
Variable part of resistor
4
Ground
Another resistor end
5
Output OFF
Short to Ground
6
N.A.
7
N.A.
8
N.A.