7.2. Manometer specification
Accuracy
±0.3% FSO (25°C)
Repeatability
±0.5% FSO
Linearity/Hysteresis
±0.29% FSO
Pressure Range
±
5 psi
Maximum Pressure
20psi
Response Time
0.5 Seconds typical
Over range Indicator
Err. 1
Under range Indicator
Err. 2
Units
Range
Resolution
PSI
5.000
0.001
Mbar
344,7
0.1
inH2O
138,3
0.1
mmHg
258,5
0.1
kPa
34,47
0,01
in Hg
10,158
0,001
Ozin
2
80,00
0,01
ftH
2
O
11,50
0,01
cmH
2
O
351,5
0,1
kgcm
2
0,351
0,001
Bar
0,344
0,001
1psi*27.68=inH2O
1psi*68.947=mbar
1psi*703.072=1*mmH2O
1psi*6894.6=Pa
FSO: Full Scale Output (The difference signal between zero and end point of range)
-17-