23. XY Format
This format is only applicable to Channel 1 and Channel 2. After the XY display format is selected, Channel 1
is displayed in the horizontal axis and Channel 2 in the vertical axis; the oscilloscope is set in the un-triggered
sample mode: the data are displayed as bright spots.
The operations of all control knobs are as follows:
* The Vertical
VOLTS/DIV
and the
VERTICAL POSITION
knobs of Channel 1 are used to set the horizontal
scale and position.
* The Vertical
VOLTS/DIV
and the
VERTICAL POSITION
knobs
of Channel 2 are used to set the vertical
scale and position continuously.
The following functions can not work in the XY Format:
* Reference or digital wave form
* Cursor
* Time base control
* Trigger control
* FFT
Operation steps:
1. Press the
Display
button and call out the
Display Set
Menu.
2. Press the
H3
menu selection button to set XY Mode
ON
. The display format is changed to be XY mode
(see
Fig. 46
).
Fig. 46
XY Display Mode