C-7100 Series User Manual PEAK INSTRUMENTS INC.
23
1) Set start wavelength and end wavelength
The start wavelength and end wavelength mean the wavelength range used for scanning.
2) Set the measurement mode
The measurement mode is absorbance, transmission or energy. Choose different measurement mode, you need to
re-set the display to display the upper and lower limits.
3) Set upper and lower limit
The different measurement modes: upper and lower display is not the same.
4) Set the time interval
Scan scanning interval setting time, 0.5 seconds minimum, 1 minute maximum.
5) Set scanning speed
The scanning speed depends on the quantity of data collecting on single point wavelength. the faster the speed, the
less the collecting quantity.
6) select to make data smoothing
The function for data smoothing is to reduce the irregular fluctuation caused by external environment during
testing.
Note: You must set the scan settings before calibration parameters blank, because the scan settings to modify the
parameters, the baseline will lead to an invalid user, the user needs to re-establish a baseline.
3. System baseline
Before the beginning of the measurement wavelength scan, you must create a system baseline. If you have
previously established a system baseline, you can skip. A long time without updating the system, re-create the
system baseline.
Press [Menu] key to view the baseline data after building system baseline.