C-7100 Series User Manual PEAK INSTRUMENTS INC.
10
accuracy, the slits there are two general representation of the width, the actual width of the slit of a knife-edge
between the two expressed (in mm), the other to represent the bandwidth of spectral bands (in nm).
D.
Spectroscopic Room: contains a half-mirror, focusing lens and reflector to achieve a beam monochromator out
into two beams of light in different directions
E.
Sample Chamber: fixed 4-cell holder. 8-cell changer is optional.
F.
Signal Board: Transfer to the motherboard processed signal detection light amplification
G.
Main board: instrument micro control unit, control instruments light source switch, the motor rotates, the signal
processor display, etc.
5. Instrument installation
1)
After unpacking, carefully check the packing list inside the object is complete and intact;
2)
Determine whether the work environment to meet the foregoing requirements, the ambient temperature 10 ~
35
℃
, relative humidity less than 85%, operating voltage (220 3 22) V / (50 3 1) H3
3)
The instrument is placed on a horizontal platform, the instrument should avoid direct sunlight and away from
electromagnetic launchers and high-power electrical devices, the use of the environment can not have dust,
corrosive gases and vibration;
4)
Around the instrument can not be any obstacle to the flow of air around the instrument;
5)
The company supplied with a power cord and make sure there are well-grounded power outlet line;
6)
Check the sample chamber, ensure that there is not any solution, and foreign matter and the process of self-test
to ensure that the sample compartment lid is closed, you can not half-open (this is very important,
Otherwise affect the instrument self-test results and normal use!!!).
7) Turn on the instrument. Then the instrument make self-test. after self-test, the instrument can be operated
normally. in case there is an error alarm halfway, please refer to the chapter of instrument troubleshooting.