PEAK
Scientific Instruments Ltd
ZA300 Zero Air Gas Generator
Instructions for use Manual
30
th
March 2005
ZA300 issue 3
7
1
Introduction
The Peak Scientific Instruments range of Zero Air Gas Generators is designed to produce a constant flow of
Zero Grade Air with a Hydrocarbon content (as Methane) of less than 0.1 ppm.
2
Unpacking and Installation.
Although Peak Scientific take every precaution with safe transit and packaging, it is advisable to fully
inspect the unit for any sign of transit damage.
ANY DAMAGE SHOULD BE REPORTED IMMEDIATELY TO THE CARRIER AND
PEAK SCIENTIFIC OR THE DISTRIBUTOR FROM WHERE THE UNIT WAS
PURCHASED.
After unpacking and a visual inspection, the unit should be placed in a ventilated area away from direct
sunlight. Care should be taken not to obstruct the ventilation holes on the sides of the unit not the fan
outlet to the rear. The generator will be warm to touch and will give off heat.
The generator should be placed on a steady and level base. It is designed to fit under most workbenches
alternatively it may be placed on a bench or in any convenient location.
Performance of the generator (like all sophisticated equipment) is affected by ambient temperatures and
humidity. Continuous operation in ambient temperatures exceeding 25
°
C will lead to a reduction in
capacity and prolonged operation in temperatures exceeding 30
°
C will shorten the life of the unit.
Operation in relative humidity exceeding 70% may overload the filtration system and result in moisture
carry-over. Additional moisture traps should be installed downstream of the generator should ambient
regularly exceed 70% Rh. and also if the generator is any distance from the application. Moisture traps or
coalescers should be located as close to the application as possible. Note should also be taken of the
proximity of Air Conditioning outlets. These can sometimes give rise to “pockets” of air with high relative
humidity. Operation of the generator within such a pocket could adversely affect its performance.
3
Electrical Connection
Important Electrical Notice
This unit is classified as SAFETY CLASS 1 equipment. THIS UNIT MUST BE EARTHED. Before
connecting the unit to the mains supply, please check the information on the serial plate. The
mains supply must be of the stated AC voltage and frequency.
EARTH/GROUND (E):-
Green & Yellow
or
Green
LIVE (L):-
Brown
Black
Neutral (N):-
Blue
White
Fuse
The generator protection fuse is found in the pull out drawer of the mains inlet IEC euro
connector located on the bottom right hand side of the cabinet adjacent to the off/on switch. The
fuse is rated at 6.3 AMP. A spare fuse is also provided in the drawer.
Connect the generator to a single phase supply using the power cord provided.