For Precision SL generators with automatic pumped water fill, ensure the water supply
bottle is not located above the height of the generator, and that the supply is not
pressurised. Generators with automatic valved water fill may be used with pressurised
deionised water supplies between 1-3 bar.
Electrical requirements are 100-240V rms nominal. However, running continuously
at voltages outwith this range is not recommended. Extended periods at extremes
can have a detrimental effect on the operation and life of the Generator.
Ensure that all pipework between generator and equipment is checked for leaks.
Please familiarise yourself with the full
User Manual
available at
www.peakscientific.
com/downloads
This includes all other technical specifications, operating
instructions, service requirements, contact details and trouble shooting.
The fixed delivery pressure of 100psi (6.9 bar) from the Precision H2 SL is suitable
for all GC-FIDs from major manufacturers. The gas pressure will be managed by
the GC. If for any reason you need to regulate the pressure to below 100psi, please
contact Peak for details of an appropriate pressure regulator.
General Notes
/
×
×
Pressurized supply is not suitable
for automatic pumped water fill.