3.1 Specification
Pressure Range
Accuracy
Over Pressure limit
Bust Pressure
Pressure Unit
Workable
Temperature Range
Storage
Temperature Range
Temperature
Compensation Range
RS232 Communication
Power Requirement
Media of Use
Display
Display Speed
Analog Output
Pressure Port
Weight
Size
Data Logging
-100 kPa to 150 MPa
0 to 250 MPa
0 to 500 MPa
Gauge : ±0.025% F.S
Absolute / Compound : ±0.05% F.S
±0.05% F.S
±0.1% F.S
** Included Nonlinearity, Hysteresis, Repeatability, errors for
-10 ~ 50 ℃ Temperature range
150 % of full scale
kPa, MPa, kg/cm
2
, psi, mbar, bar, inHg, inH
2
O, mmH
2
O, mmHg, atm
-20 ~ 70
℃
-30 ~ 80
℃
-10 ~ 50
℃
Provide Commands in manual,
Available to use private communication cable (Option)
AA Alkaline battery 3ea,
Hours of battery use approximately 1000 hours
External power supply(Option, 12~24 VDC),
Gas & Liquid (Only Gas available below 100 kPa)
5 Digits, Backlight On/Off, Auto Off
3 times / Second
(10 times automatic change if Peak function use)
1 - 5 VDC (Must use external power supply), 0.1% Accuracy
(Option : 4-20mA , 0-5VDC , 0-10VDC)
1/4" NPT M, 1/4" PF (G), 1/4" PT (R), 9/16" UNF Cone Threaded
(HF4/AF250C, 100 MPa / 150 MPa)
530 g
Dia. 110 (W) mm x 38 (D) mm x 150 (H) mm Included Pressure Port
1 time / 1, 3, 5, 30, 60 second, Store Max 3000 data
Refer to pressure type and range in catalog.