PDR1000

Digital Pressure Gauge

 

Users  Manual

 

25

Summary of Contents for PDR1000

Page 1: ...PDR1000 Digital Pressure Gauge Users Manual 25...

Page 2: ...nction Select 33 4 2 1 Alarm 33 4 2 2 Pressure Switch Test 35 4 2 3 Calibration 38 4 2 4 Baud Rate 39 4 2 5 RS232 Communication 40 4 2 6 Auto Power Off 42 4 2 7 Data Logger 43 4 2 8 Storage Data Clear...

Page 3: ...tion is used in throughout the manual to identify conditions or actions that could cause harm to the PDR1000 or user of the PDR1000 Note is used throughout the manual to identify operating and applica...

Page 4: ...recision pressure from 100 kPa to 500 MPa without a separate external power supply in anytime any place including on site and laboratory Also PDR1000 equipped solid and stylish aluminum outer casing f...

Page 5: ...Pa kg cm2 psi mbar bar inHg inH2O mmH2O mmHg atm 20 70 30 80 10 50 Provide Commands in manual Available to use private communication cable Option AA Alkaline battery 3ea Hours of battery use approxima...

Page 6: ...n and set for Alarm Switch CAL Baud Rate Auto Power Off and Data Logger Saved peak value will be clear when PDR1000 turns off Change the pressure units and On Off for Analog output In configure mode p...

Page 7: ...nter key to set up the Menu D Sub 9 Pin Map Figure 2 RS 232 Port Table 1 Pin outs 5 6 7 8 9 4 3 2 1 Pin No Description Note 1 GND 2 RX RS232 3 TX 4 Analog Out Analog Output 5 GND 6 Ext Switch In 7 Ala...

Page 8: ...Push PEAK MENU key again return to the current pressure value 4 1 3 UNITS This function can change engineering units on PDR1000 When push UNIT key engineering units will change as following steps MPa...

Page 9: ...t can select the set up when Pressure Switch mode Push Enter key for 2 seconds to start or stop for Data Logger function To set Data Logger function turn on the Data Logger function on configuration m...

Page 10: ...Scale Push ZERO key and UNITS key to change percentage value After push Enter key to save Alarm Low value Alarm Hi value setup screen shows with blink Alarm Hi value setup screen on above Digits on ab...

Page 11: ...ctric Contact Alarm Signal setup Electric Contact Alarm Signal operation mode 4 2 2 Pressure Switch Test Figure 4 Switch test Setup flowchart Normal Open N_O 5V Hi Lo 0V 0V 5V Pressure Pressure Normal...

Page 12: ...o On 2 Operation for External Pressure Switch On to Off Switch icon will change from open to close then pressure value will be hold status with Hold icon blinking Even supply pressure changed Pressure...

Page 13: ...tus then connect External Pressure Switch to PDR1000 If pressure value reach the Pressure Switch set value display pressure value will be hold then copy down the held pressure value Push ZERO key to r...

Page 14: ...be adjusted within 10 of full scale This function is Full Scale pressure value adjustment Push Enter key for 2 seconds to change Full Scale value After Full Scale adjustment finished Normal Open type...

Page 15: ...ed within 10 of full scale value and displayed full scale value adjust with Pressure controller calibrator or Piston gauge Push PEAK MENU key for 2 seconds to select Menu function Push again PEAK MENU...

Page 16: ..._No 2020G01 0001 Sensor Range 2 0 2 0 K ETX Response Example 2 Data Response Configuration 1 Byte 5 Byte 1 Byte 1 Byte 1 Byte STX ID D1 CR LF EOT ID will shows as xxx D1 has up to 25 Bytes of output d...

Page 17: ...Byte ID BL C1 BL C2 CR 001 GET MDAUTO 001 P 0 00136 T 24 95 001 P 0 00136 T 24 95 001 P 0 00132 T 24 95 001 P 0 00126 T 24 95 Response Example CR 5 Sample one Pressure value one Temperature value 5 B...

Page 18: ...hoose Auto Power Off function Push ZERO key to set the necessary interval Default Auto Power Off time set is 10 minutes Auto Power Off function can set Auto Off This function can use after Auto Power...

Page 19: ...function Push ZERO key to set Data Logger function Push ENTER key to save Data Logger status Push ENTER key to go to next Menu when Data Logger off push ENTER key to set Data Logger function save and...

Page 20: ...her various functions Data Logger function can save 100 sample data per page Data saved value are average and max Push ENTER key for 2 seconds to stop Data Logger function Available to check the data...

Page 21: ...is erase saved data on EPROM 4 2 8 End This function is save all of Menu This is last function to activate and store for each configured setting Push ENTER key to save all of function If user do not...

Page 22: ...alibration PDR1000 is reference pressure measuring gauge It needs calibration by authorized organization every year if user wants to maintain PDR1000 guaranteed accuracy Change the slope value with ca...

Page 23: ...Low Battery Change Battery Poor Power Board Contact to PDK Poor Main Board Contact to PDK Poor Display Board Contact to PDK Pressure value goes down Poor pressure tube Change tube leak check Poor Sen...

Page 24: ...ymptom except user s fault within 3 years Even when warranty is expired PDK will help our customer with minimum repair fee PDK Service Center e mail pdk pdk co kr Address Contact PDK Co Ltd 10 6 Expo...

Page 25: ...332 27 MPa 9 869 239 1 10 000 00 10 000 00 145 037 7 10 197 16 7 500 622 101 971 6 hPa 0 000 987 0 000 100 1 0 0010 00 0 014 504 0 001 020 0 750 062 10 197 16 Bar 0 986 923 0 100 000 1 000 000 1 14 50...