Your Source for Healthcare and
Wellness Infotainment
™
MODEL NUMBER:
PDI-E42LEDA-NK LED TV
Document Number:
PD196-278R3
Service Manual
Page 11 of 11
PDi Communication Systems, Inc. 40 Greenwood Lane Springboro, Ohio 45066 USA PH 1-800-628-9870 FX 937-743-5664
Safety Checks
Leakage Current
After re-assembly of the unit, always perform a leakage test on all exposed conductive parts such as metal connectors,
chassis, and mount.
With the unit powered, the following circuit should be connected from earth ground to each exposed conductive part on
the repaired unit. A DMM is placed in parallel with this circuit to measure AC rms or DC voltage. If leakage currents are
greater than the maximum allowed, inspect the TV for insulation or component breakdowns.
The maximum allowed leakage current depends on the type of power the unit receives and whether the unit is within
reach of the patient bed/chair.
TV Type
Max Leakage Current
Max DMM Voltage Reading
Footwall TV
(120VAC grounded wall outlet)
500µA AC rms
500mV AC rms
12VDC Arm Mount TV
(PDI-750XL power supply)
100µA DC
100mV DC
28VAC Arm Mount TV
(PDI-750A 750B or 772HE power
supply)
50µA AC rms
50mV AC rms
Contact Information
For any questions with servicing PDi televisions contact our Service Department at:
PDi Communication Systems, Inc.
40 Greenwood Ln.
Springboro, OH 45066 USA
800-628-9870