A4 WATER COOLING OPTION PCO.AQUAMATIC II
A4.2
FIRST TIME INSTALLATION
Take care to place the unit on a flat and firm surface. Do not cover
the cooling vents of the unit. Ensure free airflow around the
pco.aquamatic for maximum cooling performance. All tubes and
power cords need to run kink-free.
Before installation of the unit carefully read the Innovatek Protect
IP safety datasheet (see
outlet flow
inlet flow
Follow Steps 1 – 6
1.
Connect tubes to cooling unit and camera. The two arrows on the
housing of the cooling unit only indicate flow direction. Either
connection of the camera can be used for
in
or
out
.
2.
Connect to power.
3.
Open tank cover.
4.
Turn power switch on (I).
•
Slowly fill in the coolant while the unit is running. Refill as needed
to keep the level.
•
While the cooling liquid flows back to the tank make sure all air
gets out of the system – this takes a few minutes (move hoses if
necessary).
The cooling liquid tank is full when liquid level is approximately 1-2
cm from the top of the tank. The integrated pump only works when
the pump chamber is completely filled. To ensure this Move hoses or
remove air by evacuating. Tank capacity is approximately 500 ml.
After steps 1 – 6 are completed successfully the system is ready for
operation.
1
1
2
3
4
5
6
2
3
4
5
NOTE
The hose connectors are waterproof in not connected state.
Maybe they lose one drop of cooling liquid from time to time.
There is no need to empty the hoses while storing the camera
system.
6