8

 
3-16 Battery compartment/Cover 
3-17 Stand 
3-18 SD card socket 
 
 

4. Measuring Procedure 

4-1 Type K measurement 

 

1)   Power on the meter by pressing the "Power button" (3-2, Fig. 1) once. 

*  After already power on the meter, pressing the "Power button" once (> 2 sec) will turn off the 

meter. 

2)  Meter default Temp. sensor type is Type K, the display will show " K " indicator. The default 

temperature unit is Cº (Fº), the method to change the Temp. unit from Cº to Fº or  Fº please 
refer to Chapter 7-5, page 18. 

3)  Insert the Type K probes into the " T1, T2, T3, T4 input socket " ( 3-9, Fig. 1 ). The LCD will 
  show the 4 channels ( T1, T2, T3, T4 ) temperature value at the same time. 

* If the certain channels do not insert the temperature probes, the relative channel display will 

show over range " - - - - - ". 

 

4-2 Type J/T/E/R/S measurement 

All the measuring procedures are same as the Type K (section 4-1) except to select the Temp. 
Sensor type to "Type J/K/T/E/R/S" by pressing the "Type Button" (3-5, Fig. 1) once in sequence 
until the up LCD display show "J/T/E/R/S" indicator. 
 

4-3 Pt 100 ohm measurement 

1) All the  measuring  procedures  are  same  as  the  Type  K  (section  4-1)  except  to  select  the 

Temp.  Sensor  type  to "Pt"  by  pressing  the  "Type  Button"  (3-5, Fig. 1) once in sequence 
until the right down  LCD display show " Pt " text as : 
 

 

 PT 

 
 

2)  Insert the Pt 100 ohm probe ( optional, PT-101 )  to 

PT1 input socket ( 3-10, Fig. 1 ) PT2 input socket ( 3-11, Fig. 1 )

 

*  The Pt 100 ohm measurement only allow max. two channels ( two probes ) input. 
 
 

4-4 T1-T2 measurement

 

If the meter already insert two probes: Type K/J/T/E/R/S : T1, T2 input socket Pt 100 ohm : 

PT1, PT2 input socket

 

Pressing  the  "  T1-T2  button  "   (  3-6,  Fig.  1  ),  display  will show  the  difference  temperature 

value between T1, T2 (PT1, PT2) as : 

 

 

Summary of Contents for T 390

Page 1: ... Technical Services Ltd Unit N12B Tursdale Business Park Co Durham DH6 5PG United Kingdom Phone 44 0 191 377 3398 Fax 44 0 191 377 3357 info tursdaletechnicalservices co uk http www industrial needs com Manual Digital Thermometer PCE T 390 ...

Page 2: ...mpling time 0 second 10 5 4 Check time information 11 5 5 Check sampling time information 11 5 6 SD Card Data structure 11 6 Saving data from the SD card to the computer EXCEL software 11 EXCEL data screen for example 12 7 Advanced Setting 12 7 1 Set clock time Year Month Date Hour Minute Second 13 7 2 Decimal point of SD card setting 13 7 3 Auto power OFF management 13 7 4 Set beeper sound ON OFF...

Page 3: ...to 0 second during execute the manual datalogger function it can set the different position location No position 1 to position 99 Innovation and easy operation computer is not need to setup extra software after execute datalogger just take away the SD card from the meter and plug in the SD card into the computer it can down load the all the measured value with the time information year month date ...

Page 4: ...1Ԭ 1 Ԭ Datalogger Sampling Time Setting range Auto 1 second to 3600 seconds Sampling time can set to 1 second but memory data may loss Manual Push the data logger button once will save data one time Set the sampling time to 0 second Manual mode can also select the 1 to 99 position Location no Memory Card SD memory card 1 GB to 16 GB Advanced setting Set clock time Year Month Date Hour Minute Secon...

Page 5: ...optional RS232 cable UPCB 02 will get the RS232 plug Connect the optional USB cable USB 01 will get the USB plug Power off Auto shut off saves battery life or manual off by push button Operating Temperature 0 to 50 Cº Operating Humidity Less than 85 R H Power Supply Alkaline or heavy duty DC 1 5 V battery UM3 AA x 6 PCs or equivalent DC 9V adapter input AC DC power adapter is optional Power Curren...

Page 6: ...Type Resolutio n Range Accuracy Type K 0 1 Cº 50 1 to 100 0Cº 0 4 1 Cº 50 0 to 999 9 Cº 0 4 0 5 Cº 1 Cº 1000 to 1300 Cº 0 4 1 Cº 0 1 Fº 58 1 to 148 0Fº 0 4 1 8 Fº 58 0 to 999 9 Fº 0 4 1 Fº 1 Fº 1000 to 2372 Fº 0 4 2 Fº Type J 0 1 Cº 50 1 to 100 0Ԩ 0 4 1 Ԩ 50 0 to 999 9Ԩ 0 4 0 5 Ԩ 1 Cº 1000 to 1150Ԩ 0 4 1 Ԩ 0 1 Ԭ 58 1 to 148 0 Ԭ 0 4 1 8 Ԭ 58 0 to 999 9 Ԭ 0 4 1 Ԭ 1 Ԭ 1000 to 2102 Fº 0 4 2 Ԭ Type T 0...

Page 7: ...ton Button Time check n OFFSET nput socket cket Pt 10 cket Pt 10 ut terminal ter socket ut er Screws Ԩ Ԭ cy value is s cy is tested ature within ty Correction ze the therm nder the env 3 Front cklight Butto n on L Button R Button k Button Button Sa Type K T 00 ohm 00 ohm 0 to 600 Ԩ 601 to 1500 32 to 1112 1113 to 273 specified for under the m 23 5 Cº n mocouple s c vironment R t Panel on mpling tim...

Page 8: ...isplay will show over range 4 2 Type J T E R S measurement All the measuring procedures are same as the Type K section 4 1 except to select the Temp Sensor type to Type J K T E R S by pressing the Type Button 3 5 Fig 1 once in sequence until the up LCD display show J T E R S indicator 4 3 Pt 100 ohm measurement 1 All the measuring procedures are same as the Type K section 4 1 except to select the ...

Page 9: ...REC MIN symbol along with the minimum value will appear on the display If intend to delete the minimum value just press the Hold Button 3 3 Fig 1 once the display will show the REC symbol only execute the memory function continuously c To exit the memory record function just press the REC button 2 seconds at least The display will revert to the current reading 4 7 LCD Backlight ON OFF After power ...

Page 10: ...logger function stop to save the measuring data into the memory circuit temporally In the same time the text of REC will stop flashing Remark If press the Logger Button 3 8 Fig 1 once again will execute the Datalogger again the text of REC will flashing c Finish the datalogger During pause the Datalogger press the REC Button 3 4 Fig 1 continuously at least two seconds the REC indicator will be dis...

Page 11: ...mns then will generate a new file for example TMA01002 XLS 3 Under the folder TMA01 if the total files more than 99 files will generate a new route such as TMA02 4 The file s route structure TMA01 TMA01001 XLS TMA01002 XLS TMA01099 XLS TMA02 TMA02001 XLS TMA02002 XLS TMA02099 XLS TMAXX Remark XX Max value is 10 6 Saving data from the SD card to the computer EXCEL software 1 After execute the Data ...

Page 12: ... once a while in sequence to select the seven main function the lower display will show dAtE Set clock time Year Month Date Hour Minute Second dEC Set SD card Decimal character PoFF Auto power OFF management bEEP Set beeper sound ON OFF t CF Select the Temp unit to Cº or Fº SP t Set sampling time Hour Minute Second Sd F SD memory card Format Remark During execute the Advanced Setting function if p...

Page 13: ...al point for example 20 6 1000 53 Under such situation it should change the Decimal character at first When the lower display show dEC 1 Use the Button 3 5 Fig 1 or Button 3 6 Fig 1 to select the upper value to bASI or Euro bASI Use as the Decimal point with default Euro Use as the Decimal point with default 2 After select the upper text to bASI or Euro press the Enter Button 3 4 Fig 1 will save t...

Page 14: ...er value to yES or no yES Intend to format the SD memory card no Not execute the SD memory card format 2 If select the upper to yES press the Enter Button 3 4 Fig 1 once again the Display will show text yES Ent to confirm again if make sure to do the SD memory card format then press Enter Button once will format the SD memory clear all the existing data that already saving into the SD card 8 Power...

Page 15: ...ation A RS232 lead with the following connection will be required to link the instrument with the PC serial port The 16 digits data stream will be displayed in the following format D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Each digit indicates the following status D15 Start Word D14 4 D13 When send the upper display data 1 When send the lower display data 2 D12 D11 Annunciator for Disp...

Page 16: ...g screen and finish the offset adjustment procedures The above offset adjustment for Type K J T E R S is valid for T1 T2 T3 T4 at the same time 11 2 Pt 100 ohm offset adjustment 1 Set the function to Pt 100 ohm 2 Insert the Pt 100 ohm probe to the PT1 PT2 input socket 3 Pressing Offset button 3 8 Fig 1 continuously at least two seconds then release the display will show 4 If intend to make the off...

Page 17: ...00 ohm is valid for PT1 PT2 individually 12 Optional Type K Temp probe Type K TP 01 Max short tern operating Temperature 300 572 It is an ultra fast response naked bead thermocouple suitable for many general purpose application Thermocouple Probe Type K TP 02A Measure Range 50 to 900 50 to 1650 Dimension 10cm tube 3 2mm Dia Thermocouple Probe Type K TP 03 Measure Range 50 to 1200 50 to 2200 Dimens...

Page 18: ...7 4 JAPAN 3151214 TAIWAN M 358970 M 359043 CHINA ZL 2008 2 0189918 5 ZL 2008 2 0189917 0 USA Patent pending In this direction will find a vision of the measurement technique http www industrial needs com measuring instruments htm NOTE This instrument doesn t have ATEX protection so it should not be used in potentially explosive atmospheres powder flammable gases ...

Reviews: