11
5-4 Check time information
During the normal measurement (not execute the Datalogger), If press "Time check Button" (3-
7, Fig.1) once, the lower LCD display will present the time information of Year/Month,
Date/Hour, Minute/Second.
5-5 Check sampling time information
During the normal measurement ( not execute the Datalogger ), If press " Sampling Button " (
3-8, Fig. 1 ) once , the lower LCD display will present the Sampling time information in second
unit.
5-6 SD Card Data structure
1)
When the first time, the SD card is used into the meter, the SD card will generate a folder:
TMA01
2)
If the first time to execute the Datalogger, under the route TMA01\, will generate a new file
name TMA01001.XLS. After exist the Datalogger, then execute again, the data will save to
the TMA01001.XLS until Data column reach to 30,000 columns, then will generate a new
file, for example TMA01002.XLS
3)
Under the folder TMA01\, if the total files more than 99 files, will generate a new route, such
as TMA02\........
4)
The file's route structure : TMA01\
TMA01001.XLS TMA01002.XLS
TMA01099.XLS TMA02\ TMA02001.XLS
TMA02002.XLS
TMA02099.XLS TMAXX\ Remark : XX : Max. value is 10.
6. Saving data from the SD card to the
computer (EXCEL software)
1) After execute the Data Logger function, take away the SD card out from the "SD card
socket" ( 3-18, Fig. 1 ).
2) Plug in the SD card into the Computer's SD card slot ( if your computer build in this
installation ) or insert the SD card into the " SD card adapter ". then connect the " SD card
adapter " into the computer.
3) Power ON the computer and run the "EXCEL software". Down load the saving data file (for
example the file name: TMA01001.XLS, TMA01002.XLS) from the SD card to the computer.
The saving data will present into the EXCEL software screen (for example as following
EXCEL data screens), then user can use those EXCEL data to make the further Data or
Graphic analysis usefully.