5.10 Adjust the data transfer settings (RS-232)
Hold the
key when you turn on the scale, until the self-test is finished. Now the display should
show the “Baud Rate” settings. Here you can choose between 2400, 4800 and 9600 by using the
key. To confirm the selection, press the
key.
Next you can choose betwee
n “ST” and “Co”.
“ST” (stable)
The data is only transferred if the value is stabilised
“Co” (continue)
The data is transferred continuously
Press
to confirm the selection.
5.11 Calibration
You can calibrate the scale by using an external calibration weight.
To do so, hold the
key when you turn on the scale until the self-test is finished. You should now
see “CAL” on the display. Press the
key again. The display should now show “0“.
Now you can enter the weight of the external calibration weight which you want to use for the calibration
by using the numeric keys.
Note: The weight has to be entered in kg. We recommend a calibration weight which covers ca. 2/3 of the
measuring range of the scale.
Next, put the weight on the tray and press the
key to start the calibration. The calibration is now
finished.
5.12 Recharge the battery
If the battery voltage is low an indication appears on the display.
If this is the case, please turn off the scale and start charging the battery. While the battery is charged a
yellow LED is glowing. When the battery is fully charged this LED turns red.
It takes about 12 hours until the battery is fully charged.