2.10
Setting of the White Balance
Setting the white balance
Proceed as follows:
1. Attach the endoscope probe that you want to use for the inspection.
2. Switch the device on.
3. Aim the endoscope at a white surface that is within the respective probe's
working distance, but which is not more than max. 5 cm away, and bring the
image into focus.
4. Press the
White balance
button on the front of the controller.
>
The message,
White balance
, will be displayed on the screen.
Wait until the message
White balance OK
appears on the screen.
Completed!
2.11
Setting of the Light Amplification on the Screen
Setting the light amplification
Adapt the screen image to a dark environment by intensifying the monitor display.
Note: You will not notice any difference when adjusting the setting in a bright or
highly reflective environment.
Proceed as follows:
1. Press the
GAIN
button on the front of the controller.
>
The display on the screen will intensify.
>
The current setting will display on the screen ("ACG 08 dB").
>
The LED light amplification that you have set will remain unchanged.
2. Keep pressing the
GAIN
to continue amplifying the amount of light shown on
the screen.
>
The current setting will display on the screen ("ACG 12 dB").
Pressing the button one more time will set your light amplification on the
screen back to 0 dB.
Completed!