21
Warning
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery
as soon as the battery indicator “ ” appears.
Make sure the test leads are disconnected from the circuit being tested before opening the case bottom.
Follow Figure 5-2 and proceed as follows to replace the battery:
z
Turn the rotary switch to OFF and remove all connections from the terminals.
z
Remove the screw from the battery compartment, and separate the battery compartment from the
case bottom.
z
Replace with 4 pieces new 1.5V
(R6)
batteries.
z
Rejoin the case bottom and battery compartment, and reinstall the screw.
Specifications
Safety and Compliances
Maximum Voltage between any
Terminal and Grounding
Refer to different range input
protection voltage
Certification
Compliances IEC
61010
CAT.
II
1000V,
CAT.
III
600V overvoltage
and double insulation
standard
Fused Protection for
mA
input terminal:
0.5A, 250V, fast type fuse,
ø5×20mm
Fused Protection for
10A
input terminal:
10A , 250V,fast type fuse,
ø5×20mm
Physical Specifications
Display (LCD)
Digital: 3999 counts on display;
updates 2-3
times / second.
Operating
Temperature
0
Ԩ
~40
Ԩ
(32
Ԭ
~104
Ԭ
)
Storage
Temperature
-10
Ԩ
~50
Ԩ
(14
Ԭ
~122
Ԭ
)
Relative Humidity
≤
75%
@
0
Ԩ
~40
Ԩ
;
≤
70%
@
-10
Ԩ
~50
Ԩ
.
Altitude Operating:
2000m;
Storage: 10000m.
Battery Type
1.5V (R6) x 4 Batteries or Power adaptor.
Check carefully about the working voltage
of power adaptor before use.
Electromagnetic
Compatibility
z
In a radio field of 1 V/m below:
Overall Accuracy = Specified
Ac 5% of Range
z
In a radio field of 1 V/m above:
No assigned accuracy is specified.
Dimensions (H x W x
L)
200 x 100 x 48mm.
Weight
Approx. 498g (including battery)