Measurement 

Apply the TT-Gel evenly to the surface to be measured. Place the probe onto the surface covered 
with the coupling agent. 

 

7.1 

Clean surface 

Before the measurement, the surface of the object to be tested should be cleaned of 
dust, dirt and rust. Coatings such as paint must also be removed. 

 

7.2 

Reducing the roughness 

Rough  surfaces  can  cause  measurement  errors  or  missing  readings.  Before  the 
measurement, the material surface should be made as smooth as possible by grinding,  
polishing or using a high-viscosity coupling agent. 

 

7.3 

Rough surfaces due to heavy machines 

Regular fine grooves which occur, for example,  during production  with  lathe machines  
and planers can also cause measurement errors. The corrective measure is the same as 
in chapter 7.2. In addition, a better result can be obtained by adjusting the angle using 
the silver marking on the centre of the probe head and the fine grooves of the material to 
be tested (orthogonal or parallel to the marking). 

 

7.4 

Measurement of a round surface 

To measure a round surface such as a pipe or an oil barrel, it is important to set the angle 
between the silver line of the probe and the axis of the material to be tested. In short, the 
mark in the sensor head must be kept parallel or perpendicular to the axis of the material 
under  test.  Slowly  move  the  probe  head  perpendicular  to  the  axis  across  the  material 
under  test  and  the  values  on  the  display  will  change  regularly.  The  lowest  reading 
displayed is the minimum thickness of the material being measured. 
The direction in which the probe is guided depends on the curvature of the material. For 
large diameter pipes, the mark in the probe  head should be  perpendicular to  the pipe 
axis;  for  smaller  diameter  pipes,  the  mark  in  the  probe  head  can  be  parallel  or 
perpendicular to the material axis. The lowest reading is saved as the measured value. 

 

7.5 

Measurement of composite material 

When  measuring  composite  shapes  (such  as  pipe  elbows),  the  method  described  in 
chapter 7.4 can be used. However, the measurement must be carried out twice to obtain 
two readings. The silver line of the probe must be vertical or parallel to the axis. The lower 
value is used as the measured thickness. 

 

7.6 

Non-parallel surfaces 

To obtain a satisfactory result, the surface of the material to be tested must be parallel or 
coaxial  to  the  surface  of  the  sensor,  otherwise  measurement  errors  will  occur  or  no 
reading will be obtained. 

E

n

g

li

s

h

 

Summary of Contents for PCE-TG 150

Page 1: ...nstruments User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Kal nl k l er PCE TG 75 PCE...

Page 2: ...amalar 4 3 2 Cihaz 5 3 3 Aray z 6 3 4 Ekran l m durumunda 7 4 Haz rl k 7 4 1 G Kayna 7 4 2 lk Kullan m 7 5 Men 7 5 1 l m 7 5 2 Sens r sadece PCE TG 150 7 5 3 S n r 8 5 4 Haf za 8 5 5 Analiz 8 5 6 Ayar...

Page 3: ...0 7 7 Malzeme S cakl Etkileri 10 7 8 S n mleme Malzemeleri 10 7 9 Referans rnekleri 10 8 Kalibrasyon Cal 10 8 1 Kalibrasyon Talimatlar 10 8 2 Ses H z Kalibrasyonu V Cal 11 9 Bak m 11 9 1 Temizleme 11...

Page 4: ...e 18 3 3 Interfaces 19 3 4 Display in measuring condition 19 4 Getting started 20 4 1 Power supply 20 4 2 First use 20 5 Menu 20 5 1 Measure 20 5 2 Probe PCE TG 150 only 20 5 3 Limit 20 5 4 Memory 20...

Page 5: ...faces 22 7 7 Effects of the material temperature 23 7 8 Damping materials 23 7 9 Reference examples 23 8 Calibration Cal 23 8 1 Calibration instructions 23 8 2 Sound velocity calibration V Cal 23 9 Ma...

Page 6: ...kalifiye PCE Teknik Cihazlar personeli taraf ndan a lmal d r Cihaz slak elle asla kullanmay n Cihazda herhangi bir teknik de i iklik yapmay n Cihaz sadece bez ile temizlenmelidir Sadece pH n tr temiz...

Page 7: ...limat eri i Model Teslimat eri i PCE TG 75 1 x Kal nl k l er PCE TG 75 1 x Sens r PCE TG 5M10d 1 x Temas Jeli TT JEL 3 x 1 5 V AA Pil 1 x Ta ma antas 1 x Kullan m K lavuzu PCE TG 150 1 x Kal nl k l er...

Page 8: ...zemeler i in standart sens r 10 mm yar apl borular Frekans 5 MHz ap 10 mm S cakl k Aral 10 60 C PCE TG 5M6d Uygulama D z malzemeler i in standart sens r 6 mm yar apl borular Frekans 5 MHz ap 6 mm S ca...

Page 9: ...190 x 120 x 70 mm siyah mavi 3 Sistem A klamas Kal nl k l er PCE TG 75 PCE TG 150 ultrasonik dalgalar l er ve b ylece malzemelerin kal nl tahribats z ve hassas bir ekilde belirlenebilir Cihaz retim ek...

Page 10: ...PCE Teknik Cihazlar 5 3 2 Cihaz No Tu A klama Fonksiyon 1 Sens r soketi 2 Yukar Yukar gitme ve de er artt rma 3 Geri ptal geri k 4 OK Onayla 5 Sens r ba l T rk e...

Page 11: ...ltma 7 Kalibrasyon blo u 8 A ma Kapama Men Cihaz a mak i in bas n ve b rak n Cihaz kapatmak i in bas l tutun Men y a mak i in bas n ve b rak n 9 De i tir Farkl malzemeler aras nda ge i yapma 10 Ekran...

Page 12: ...ullan m Cihaz a madan nce sens r soketine sens r yerle tirin Cihaz al t rmak i in ekranda cihaz n ad g r nt leninceye kadar a ma kapama men tu una bas n Sonras nda ses h z l m i in aray z a lacakt r M...

Page 13: ...ik halinde g r nt leyebilirsiniz a Veri numaras maksimum de er minimum de er ve ortalama de er gibi temel bilgileri inceleyebilirsiniz b Grafi i b y tebilirsiniz c Belirli bir veri de erini g r nt lem...

Page 14: ...ya makineleri ile retim s ras nda olu an ince oluklar da l m hatalar na neden olabilir Bunun d zeltici nlemi b l m 7 2 deki ile ayn d r Bunun haricinde sens r ba l n n ortas nda bulunan g m izgi ve te...

Page 15: ...rleri ger ek verilerden daha d k olur 7 8 S n mleme Malzemeleri Lifli g zenekli ve kaba par ac klar olan malzemeler ultrasonik dalgalar n g l bir ekilde sa lmas na ve enerji zay flamas na neden olarak...

Page 16: ...n kontrol l m n tekrarlay n Pas lekeleri ve korozyon l lecek y zeydeki pas lekeleri korozyon ukurlar l m de erlerinin d zensiz bir ekilde de i mesine ya da hi de i memesine neden olur K k pas lekeleri...

Page 17: ...dede farkl oldu u i in bu da hatalara neden olur Farkl oksit katmanlar kullan c n n dikkat etmesi gereken farkl hatalara neden olur Cihaz kalibre etmek i in cihaz ile kal nl kontrol etti iniz ayn malz...

Page 18: ...da bilgilendirmek zorunday z Kullan lm piller evsel at k olarak at lmamal d r T keticiler pilleri t ketici i in cretsiz olan uygun bir ticari veya belediye toplama noktas na vermekle y k ml d r Kullan...

Page 19: ...The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only b...

Page 20: ...Weight 320 g 0 70 lbs H refers to the material thickness of the sample 2 2 Delivery contents Model Delivery contents PCE TG 75 1 x thickness gauge PCE TG 75 1 x coupling gel TT GEL 3 x 1 5 V AA batte...

Page 21: ...iameter 10 mm 0 29 Temperature range 10 60 C 14 140 F PCE TG 5M6d Application Standard probe for flat materials pipes with a radius 6 mm Frequency 5 MHz Diameter 6 mm 0 24 Temperature range 10 60 C 14...

Page 22: ...48 x 4 72 x 2 76 black blue 3 System description The PCE TG 75 is a thickness gauge that measures the speed of ultrasonic waves and can thus determine the thickness of materials accurately and non des...

Page 23: ...PCE Teknik Cihazlar 18 3 2 Device No Key Description Functions 1 Probe socket 2 Up Navigate up increase value 3 Back Cancel back exit 4 OK Confirm 5 Probe head English...

Page 24: ...meter Press and release to open the menu 9 Change Change between different materials 10 Display 11 Battery compartment 3 3 Interfaces Probe socket 3 4 Display in measuring condition 1 Title bar 2 Bat...

Page 25: ...nt can be started here 5 Menu The menu of the meter can be reached by pressingthe menu key after start up With the arrow keys you can navigate through the menu You can confirm your selection with the...

Page 26: ...ersion number g Factory reset 6 Submenu 6 1 Operation a Enter the submenu Press the back key in the measuring interface to open the submenu b Select menu item Press the up and down keys to change betw...

Page 27: ...between the silver line of the probe and the axis of the material to be tested In short the mark in the sensor head must be kept parallel or perpendicular to the axis of the material under test Slowl...

Page 28: ...city is used to calibrate the meter At least one reference sample is required to calibrate the instrument The ultrasonic thickness gauge is equipped with a calibration block with a thickness of 4 0 mm...

Page 29: ...f the probe at different angles for several tests Incorrect material identification If you have calibrated the instrument with one material and use it to test another material an incorrect result will...

Page 30: ...d applied evenly The coupling agent is usually applied to the surface of the material under test If the temperature is too high the coupling agent is applied to the bottom of the probe Only PCE Instru...

Page 31: ...nit 11 Southpoint Business Park Via Pesciatina 878 B Interno 6 Halkal Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece stanbul United...

Page 32: ...ar 27 Specifications are subject to change without notice User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www p...

Reviews: