- 11 -
Operation method of re-sampling: Press “CANCEL”, back to “Entering sampling”
screen. Press “Enter” to sample. Then press “Testing”, the data will be for new
sample surface.
6.7 Testing interface
Let the instrument aim at tested sample, press
“
Testing”, the difference between
tested surface and standard sample will be displayed in main screen.
The instrument can make a primary estimation for testing data and show estimated
result for reference. the test data
△
E
、△
L
、△
a and
△
b respectively are displayed
and the state of the tested data will be shown.
The result of
△
Eab is judged by tolerance setting of system setting will be shown
in status bar. Please note the proper judging range while setting the tolerance.
Under default, the tolerance range of color difference
△
Eab is
“
1
”
Tolerance range can be set if required. Please refer to function setting of Tolerance
setting.
After accepted tolerance setting, it will give judge automatically for each test. There
will be one buzzing if the result is within the accepted tolerance, or twice buzzing
will be happened.