- 8 -
Please select “Enter to Main” to get to the sampling screen After above steps.
6.4 Correct white
ⅰ
.Correcting white calibrate of XYZ in lower machine (
PCE-
TCR
200)
Put the testing aperture on the white tabula while selecting “Correct white” and
press “enter”, then “correct white page” will be shown as follows:
The displayed values are real XYZ values of the white tabula. For changing XYZ
values, please press “enter” and change the value by “up” and “down”. Then press
“YES” to confirm the change.
ⅱ
. Correcting white calibrate of XYZ in upper monitor (Computer)
(a) Connect instrument with pc by USB communication, then open software of
upper monitor of
PCE-
TCR
200.
(b) The window “correct white” will pop up by Clicking “control” in status bar.