c) To exit the memory record function, press and hold the "REC" button for at least 2 seconds. The 
display will revert to the current reading. 
 

LCD Backlight ON/OFF 

After turning the power ON, the "LCD Backlight" will be activated automatically. During the measurement, 
press the "Backlight Button" (4-2) once to turn OFF the "LCD Backlight". Press the "Backlight Button" 
once again to turn ON the"LCD Backlight" again. 

Datalogger 

 

Preparation before using the datalogger function 

a. Insert an SD card (1 GB to 16 GB, optional) into the "SD card socket" (4-18). The front side of the SD 
card should face the bottom of the case. 
b. If you use the SD card with the meter for the first time, it is recommended to take care of the "SD card 
Format" first. 
c. Time setting 
When the meter is used for the first time, the clock time should be set exactly. 
d. Decimal format setting 
The numerical data structure of the SD card is decimal by default, for example "20.6" "1000.53". 
However, in certain countries (Europe ...) the "," is used as the decimal point, for example " 20, 6 " 
"1000,53". Under these circumstances, the decimal character should be changed first. 

 
Auto datalogger (Set sampling time 

1 second) 

a. Start the datalogger 
Press the "REC Button (4-4) once. The LCD will show the text "REC". Then press the "Logger Button" (4-
8); " REC " will flash and the beeper will sound. At the same time, the measuring data along with the time 
information will be saved to the memory circuit. 
b. Pause the datalogger 
When using the datalogger function, when you press the "Logger Button" (4-8) once, the datalogger 
function will be paused ( stop to save the measuring data into the memory circuit temporarily). "REC" will 
stop flashing. 
Note: If you press the "Logger Button" (4-8) once again, the datalogger function will be carried out again 
and the text "REC" will flash. 
c. Finish the datalogger 
When pausing the datalogger, press and hold the "REC Button" (4-4) for at least two seconds, the "REC" 
indication will disappear and the datalogger function will be finished. 
 

Manual datalogger (Set sampling time = 0 seconds) 
 

a. Set sampling time to 0 seconds 
Press the "REC Button

” (4-4) once. The LCD will show the text "REC", then press the "Logger Button" (4-

8) once, "REC" will flash once and the beeper will sound once. At the same time, the measuring data 
along with the time information and the position no. will be saved to the memory circuit. 
Note: 

* For the 4-channel measurement, the lower right display area will show the position/location no. 
(P1, P2... P99) and the T4 measurement value alternately. 
* During execution of the manual datalogger 

function, use the "▲ Button" (4-5) or the "▼ Button" 

(4-6) to set the measuring position (1 to 99, for example room 1 to room 99) to identify the 
measurement location. 

b. Finish the Datalogger 
Press the "REC Button" (4-4) continuously for at least two seconds. The "REC" indication will disappear 
and the datalogger will be finished. 

Check time information 

During normal measurement (no datalogger), when you press the "Time check Button" (4- 
7) once, the lower LCD display will present the time information Year/Month, Date/Hour, Minute/Second. 
 

Check sampling time information 

During normal measurement (no datalogger), when you press the "Sampling Button" (4-8) once , the 
lower LCD display will present the sampling time information in the second unit. 
 

Summary of Contents for PCE-T 390

Page 1: ...ts UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce instru...

Page 2: ...alogger 9 7 Saving data from the SD card to the computer EXCEL software 10 8 Advanced settings 10 9 Power supply from DC adaptor 11 10 Battery replacement 11 11 RS232 PC serial interface 11 12 Offset...

Page 3: ...ssible Data hold records max and min reading High accuracy Power by UM3 AA 1 5 V x 6 batteries or DC 9V adapter RS232 USB PC COMPUTER interface 2 Safety notes Please read this manual carefully and com...

Page 4: ...1 C 0 1 F 1 C Datalogger Sampling time Setting range Auto 1 second to 3600 seconds Sampling time can be set to 1 second but memory data may get lost Manual Pushing the datalogger button once will save...

Page 5: ...r manual off by pushing button Operating temperature 0 50 C Operating humidity Less than 85 RH Power supply Alkaline or heavy duty DC 1 5 V battery UM3 AA x 6 pcs or equivalent DC 9V adaptor input AC...

Page 6: ...0 2102 F 0 4 2 F T type 0 1 C 50 1 100 0 C 0 4 1 C 50 0 400 0 C 0 4 0 5 C 0 1 F 58 1 148 0 F 0 4 1 8 F 58 0 752 0 F 0 4 1 F E type 0 1 C 50 1 100 0 C 0 4 1 C 50 0 900 0 C 0 4 0 5 C 0 1 F 58 1 148 0 F...

Page 7: ...on button R button 4 7 SET button time check button 4 8 Logger button OFFSET button sampling time check button 4 9 T1 T2 T3 T4 input socket type K type J 4 10 PT1 input socket Pt 100 Ohm 4 11 PT2 inpu...

Page 8: ...t 100 Ohm measurement only allows max two channels two probes to be used T1 T2 measurement If two probes are inserted type K J T E R S use T1 T2 input socket Pt 100 Ohm use PT1 PT2 input socket When y...

Page 9: ...the datalogger function will be paused stop to save the measuring data into the memory circuit temporarily REC will stop flashing Note If you press the Logger Button 4 8 once again the datalogger fun...

Page 10: ...al times to select one of the seven main functions The lower display will show dAtE Set clock time year month date hour minute second dEC Set SD card decimal character PoFF Auto power OFF management b...

Page 11: ...utton 4 5 or the Button 4 6 to set the upper value to yES or no yES to format the SD memory card no not format the SD memory card 2 If you select yES press the Enter Button 4 4 once again The display...

Page 12: ...egative D9 Decimal point DP position from right to left 0 no DP 1 1 DP 2 2 DP 3 3 DP D8 to D1 Display reading D1 LSD D8 MSD e g if the display reading is 1234 then D8 to D1 is 00001234 D0 End Word RS2...

Page 13: ...least two seconds then release it The display will show 4 If you wish to make the offset adjustment for Pt 1 insert the probe into the PT1 input socket Press the L button 4 5 once so that the display...

Page 14: ...0 cm tube 3 2 mm Thermocouple probe Type K TP 03 Meas range 50 1200 C 50 2200 F Dimensions 10 cm tube 8 mm Surface probe Type K TP 04 Meas range 50 400 C 50 752 F Dimensions temp sensing head 15 mm Pr...

Page 15: ...contact PCE Instruments 15 1 PCE Instruments UK By post PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign Way Southampton Hampshire United Kingdom SO31 4RF By phone 02380 987 035 15 2...

Reviews: