Introduction 

 

Thank you for selecting the PCE-PWT 10 pocket salinity tester. This manual provides a step-by-

step guide to help you operate the meter, please carefully read the following instructions before 

use. The following list describes the standard accessories of the tester. After unpacking, please 

check  that  all  accessories  are  complete.  If  any  are  damaged  or  missing,  please  contact  the 

nearest distributor. 

 

Keypad 

 

PCE-PWT 10 pocket salinity tester has a succinct membrane keypad, names and symbols which 

describe each of the function key controls. 

 

KEY  

FUNCTION  

MEAS/HOLD  

 Power the tester ON/OFF.  

 Freezes the measured values on the 

display, press the key again to resume 

measuring.  

 In the calibration mode, exits calibration 

and returns to measurement mode.  

 

 Press the key to enter the calibration 

mode.  

 Press and hold the key to enter setup 

menu.  

 In the calibration mode, press the key to 

set calibration values.  

 In the setup mode, press the key to select 

default option.  

ENTER  

 Confirms the calibration or displayed 

option.  

 Toggles between conductivity and salinity 

measurement modes.  

 

 

 

Summary of Contents for PCE-PWT 10

Page 1: ...ter Analysis Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last cha...

Page 2: ...int Calibration 8 10 2 Point Calibration 9 11 3 Point Calibration 10 12 Temperature Calibration 10 13 Switching the Measurement Mode 11 14 Measurement 11 15 Hold Function 11 16 Electrode Care and Main...

Page 3: ...ons The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should on...

Page 4: ...0 pocket salinity tester has a succinct membrane keypad names and symbols which describe each of the function key controls KEY FUNCTION MEAS HOLD Power the tester ON OFF Freezes the measured values on...

Page 5: ...er is in the measurement mode CAL Indicates tester is in the calibration mode SETUP Indicates tester is in SETUP mode HOLD Indicates the measured value has been frozen ATC Indicates temperature compen...

Page 6: ...ff to conserve energy To disable the auto off function please read the SETUP MENU section PARAMETER DESCRIPTION OPTION S DESCRIPTION DEFAULT Calibration Point 1 point 2 points 3 points Temperature Uni...

Page 7: ...tester goes to temperature unit selection mode the display shows o C UNIT 3 Press CAL key to select the temperature unit o C or o F and press ENTER key to confirm The tester goes to temperature calibr...

Page 8: ...function is enabled When the auto off function is enabled if you do not press any keys for 8 minutes the tester will automatically turn off to conserve energy 7 Press CAL key to enable or disable the...

Page 9: ...te the meter When the calibration is done all new calibration values will automatically override existing data The following table shows acceptable conductivity range of calibration solution for each...

Page 10: ...immediately shows current calibration standard e g 1413 S cm 1 4 Press ENTER key the default calibration value begins flashing 1 5 Press CAL key to set each digit press ENTER key to confirm When the...

Page 11: ...de into the calibration solution the tester automatically shows current calibration standard e g 12 88mS cm 4 Press ENTER key the default calibration value begins flashing 5 Press CAL key to set each...

Page 12: ...12 Temperature Calibration During the measurement if the temperature reading displayed differs from that of an accurate thermometer you need to calibrate the tester 1 Press and hold the CAL key for 3...

Page 13: ...two data hold modes When the Auto Hold function is enabled the tester will automatically sense a stable endpoint reading and freeze it HOLD indicator appears on the display If the Auto Hold function...

Page 14: ...tly push the module into the tester Twist the electrode collar clockwise until it is tight Installation is complete 18 Replacing the Batteries If the battery indicator is disappearing during use you w...

Page 15: ...0 00 10 00ppt Accuracy 1 F S Resolution 0 01ppt Conductivity Range 0 20 00 200 0 2000 S cm 20 00mS cm Accuracy 1 F S Resolution 0 01 0 1 1 Calibration Points 1 to 3 points Calibration Solutions 10 S...

Page 16: ...neral Business Terms which you can find here https www pce instruments com english terms 24 Disposal For the disposal of batteries in the EU the 2006 66 EC directive of the European Parliament applies...

Page 17: ...583 975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english w...

Reviews: