19 Troubleshooting
LCD
DISPLAY
CAUSE
CORRECTIVE ACTION
Electrode dried out
Soak the conductivity electrode in tap water for 10
minutes
Measured value is out of
range
Check the electrode whether clogged, dirty or
broken
Incorrect calibration solutions
Using the fresh conductivity standard solutions for
calibration
Setting value does not match
calibration solution
Reset the calibration value
Electrode is broken
Replace the electrode module
20 Specifications
Salinity
Model
PCE-PWT 10
Range
0.00~10.00ppt
Accuracy
± 1% F.S
Resolution
0.01ppt
Conductivity
Range
0~20.00, 200.0, 2000µS/cm,
20.00mS/cm
Accuracy
± 1% F.S
Resolution
0.01, 0.1, 1
Calibration Points
1 to 3 points
Calibration Solutions
10µS/cm,
84µS/cm,
1413µS/cm, 12.88mS/cm
Temperature
Range
0~60°C, 32~140°F
Accuracy
± 1°C
Resolution
0.1°C, 0.1°F
Calibration Points
1 point
Calibration Range
Measured value ± 10°C
Others
Temperature Compensation
0~60°C, 32~140°F
Temperature Coefficient
2%/°C
Cell Constant
K=1
Normalization Temperature
25°C, 77°F
Hold Function
Manual or Automatic
Power Off
Manual or Automatic (8 minutes
after last key pressed)
Sensor Type
Order Code: ECSCAN-C1-10K
Power Requirements
2× 1.5V "AAA" Batteries
Dimensions
185(L)× 40(Dia.)mm
Weight
100g