0
rated operating conditions:
- supply
9...24 V a.c./d.c.
- power consumption
< 0.5 VA
- ambient temperature
- 20...23...60
o
C
- relative air humidity
< 95%
3)
- rate of air flow
0,5 m/s
4)
- preheating time
15 minutes
- protection grade ensured
by the housing
IP 65
- fixing
on a wall
- weight
125 g
- dimensions
(35
´
58
´
118) mm
- working position:
in applications non-exposed to a direct contact with water: any
in applications exposed to a direct contact with water:
with the sensor chamber directed towards the ground
electromagnetic compatibility:
- noise immunity
acc. to EN 61000-6-2
- noise emission
acc. to EN 61000-6-4
safety requirements acc. to en 61010-1
- installation category
III
- pollution grade
2
- phase-to-earth operating voltage
50 V
- altitude above sea level
< 2000 m
•
•
1)
In case of condensation of water vapor on the sensor surface,
the measurement error may exceed the basic error till a moment
of drying up the sensor structure
2)
The absolute temperature measurement range is –30...85
o
C but the
measurement class is not guaranteed beyond the basic range
3)
Admissible condensation of water vapor when using additional sensor
shields, see Tab. 9)
4)
For air flow < 0.5 ms, the temperature and humidity measurement
error may increase by 100%.