6.6 registers for readout
Value set in the two following 16-bit registers.
Registers contain the same data as 32-bit registers
from the area 7500.
The value is located in the
32-bit registers
Name
W
rite (w)/ readout (r)
Unit
Quantity name
7000
7500
ID
r
-
P18(D) device identifier
7002
7501
T
r
o
C
Measured temperature
7004
7502
RH
r
%
Measured relative humidity
7006
7503
DP
r
o
C
Calculated dew-point
7008
7504
AH
r
g/m
3
Calculated absolute humi-
dity
7010
7505
min T
r
o
C
Min. of temperature
7012
7506
max T
r
o
C
Max. of temperature
7014
7507
min RH
r
%
Min. of relative humidity
7016
7508
max RH
r
%
Max. of relative humidity
7018
7509
min DP
r
o
C
Min. of dew-point
Table 15