background image

 

 

©  PCE Instruments 

 

6.1 

Air flow measurement and temperature 

During the airflow measurement, you can also read out the current temperature in the display. 

The temperature will be shown in the bottom left-hand part of the display. 

Now press the Unit/Zero Button to switch the airflow units. In order to change the temperature, 

press the Temp./Lux Button (

/

/▲/Lux/Ft-cd). 

6.2 

Volume flow measurement 

For volume flow measurement, you can select the unit “CMM” for cubic meters or “CFM” for 

cubic feet per minute. Press the "Unit/Zero Button" to switch the volume flow unit. 

6.2.1 

Set measuring area for volume flow 

To  set  the  measuring  area  for  the  volume  flow,  first  select  the  desired  unit.  The  press  the 

Temp./Lux Button (

/

/▲/Lux/Ft-cd). The display will show “m-2” for square metres and “f-2” for 

square  feet.  Now  set  the  area.  The  Unit/Zero  Button  is  used  to  get  to  the  next  digit  and  the 

Temp./Lux  Button  (

/

/▲/Lux/Ft-cd)  and  the  Function/▼  Button  are  used  to  increase  and 

decrease  the  value  respectively.  To  confirm  the  set  value,  push  the  MIN/MAX  Button.  The 

measurement will start after a few minutes. 

 

Note

: It is not possible to enter round measuring areas. 

6.3 

Barometric pressure measurement (abs. pressure) 

To change the unit for the air pressure measurement, press the Unit/Zero Button. You can select 

HPA, inHP or mmHP. 

6.4 

K-type thermocouple 

First plug a K-type thermocouple into the socket of the device. The temperature value will be 

displayed immediately. To change the temperature unit, push the Temp./Lux Button 

(

/

/▲/Lux/Ft-cd). 

6.5 

Light measurement 

When you make a light measurement, the display will turn by 180 °. Turn the device by 180 ° as 

well to make sure the light sensor lies bare and you can read out the measurement values. Press 

the Temp./Lux Button (

/

/▲/Lux/Ft-cd) to select measuring unit "Lux" or "Ft-cd".  

6.5.1 

Zero point of the light sensor 

When the environmental temperature or the battery capacity changes, the zero point of the light 

sensor can drift. To get the most accurate measurement results, it is recommended to zero the 

device regularly. To do so, cover the sensor so that it is no longer exposed to any light. Then 

press the Unit/Zero Button to zero the device. Zeroing is only possible if the indicated value is 

below 20 lux. 

6.6 

Humidity and ambient temperature 

While measuring the current air humidity, you can also read out the current ambient temperature. 

To change the temperature unit, press the Temp./Lux Button (

/

/▲/Lux/Ft-cd). 

6.7 

Dew point measurement 

When making a dew point measurement, you can change the unit by pressing the Temp./Lux 

Button (

/

/▲/Lux/Ft-cd). 

 

Summary of Contents for PCE-EM 888

Page 1: ...88 Lux Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change 10...

Page 2: ...4 6 1 Air flow measurement and temperature 5 6 2 Volume flow measurement 5 6 3 Barometric pressure measurement abs pressure 5 6 4 K type thermocouple 5 6 5 Light measurement 5 6 6 Humidity and ambien...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...er 2 1 Delivery contents 1 x PCE EM 888 hygrometer 1 x Type K thermocouple sensor TF 500 1 x 9V battery 1 x User manual 3 Specifications Wind speed Measuring range 80 5910 ft min 0 4 30 m s 1 4 108 km...

Page 5: ...hPa 1 hPa 0 1 mmHg 0 01 inHg Accuracy 1 5 hPa 2 hPa 1 2 mmHg 0 05 inHg General specifications Display 8 mm 0 32 LCD Measurement functions Air speed velocity Volumetric air flow Relative humidity Air...

Page 6: ...AX Min recording has been started REC icon in the display you will not be able to turn off the instrument 6 Measuring Functions The meter has these 7 functions Display indication Measuring function An...

Page 7: ...ressure To change the unit for the air pressure measurement press the Unit Zero Button You can select HPA inHP or mmHP 6 4 K type thermocouple First plug a K type thermocouple into the socket of the d...

Page 8: ...old the MAX MIN Button for three seconds 9 Auto Power Off In order to prolong the battery life the instrument has an Auto Power Off function The meter will switch off automatically if no buttons are p...

Page 9: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 10: ...583 975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english w...

Reviews: